Motor Protection; Error Codes - IKA MTS 2/4 digital Operating Instructions Manual

Table of Contents

Advertisement

Available languages

Available languages

GlobalScientificSupply
www.
While Timer mode is in operation, you can increase or decrease the time
setting as required by pressing the "1" key (6) and/or the "10" key (5)
as described above. After you have changed the time setting, the time
begins to elapse again and the LED display (2) shows the remaining time
in minutes as it did before.

Motor protection

Blocking or overloading the motor beyond the permissible motor tem-
perature causes the device to switch off automatically due to the safety
circuit. The LED display (2) displays the corresponding error message. To
eliminate the error, please check the operating conditions. Also allow
the device to cool off.
The device must be turned off and back on again.

Error codes

Error message MTS 2/4:
Please note! If an error occurs, an audio warning signal is generated and
the Timer display switches to E + error number. First try to turn off the
device and then turn it back on to continue operation. If an error is not
eliminated even after waiting for a considerable amount of time, please
contact our service department.
You should always let us know what error code was displayed.
.com
This facilitates troubleshooting and helps us draw some initial conclu-
sions.
Error message
E 1
E 2
E 3
E 4
E 5
E 7
E 8
GLOBAL SCIENTIFIC SUPPLY
The Laboratory Supply Company!
Error
Cause of error
The potentiometer for the -Too rapida change in the
speed setting does not
speed target value
specify any target value
-Internal device error
Motor blocked
-Agitation table is inhibi-
ted in its stroke motion by
an external effect
Read fork signal not
-Internal device error
present
Maximum adjustable
-Resonance behavior
speed of the device
of the test setup
exceeded
Motor standstill for de-
-To switching on the
vice switch-on routine
device, the motor does
(safety query)
not en rotation motion
not provided
Safety relay cannot be
-Internal device error
activated
Problem with power
-The device is being ope-
supply
rated with over-voltage or
under-voltage. (For per-
missible voltage range,
see the Technical Data)
Triac cannot be activated
-Internal device error
info@GlobalScientificSupply.com