The Function Test Is Successfully Completed When; A Successful Test Will Be Confirmed By A Green, Ticked Circle.; Use Irwin App To Transfer Your Saved Log Files To Your Computer - Inficon IRwin Methan Leak Detector Quick Start Manual

Table of Contents

Advertisement

11

The Function Test is successfully completed when

the graph shows detection of methane, ethane and
propane
12
A successful test will be confirmed by a green, ticked
circle. You can also save the GC report.
13
Use IRwin App to transfer your saved log files to your
computer and create pdf file of the test.
IRwin App is available on www.inficon.com
How to Perform GC Function Test on IRwin Methane Leak Detector
ww w. i nf i co n .com
r ea c h . s w e d e n @ i n f i c o n . c o m
Due to our continuing program of product improvements, specifications are subject to change without notice.
All trademarks are the property of the ir respective owners.
©2019 INFICON
mimd66en1-03 (1910)
IRwin SXG
S/N: 92001853
S/W ver: 3.06.01
10/10/2019
Function Test Result GC
13:46:53
60s
120s
180s
Gas Concentration 1.6 %
CH4 (Methane)
C2H6 (Ethane)
C3H8 (Propane)