19. GENERAL SPECIFICATION
Max Intensity
50V zero to peak. Setting 0-100 in steps of 0.5
OC cutout below 160 Ohm. Constant current 160-470 Ohm 470-
2000 Ohm constant voltage
Channels
Single
Waveform
Asymmetrical rectangular
Max Pulse energy
* Total output limited to 25uC per pulse
Power
BL-4B Li-ion battery 3.7V
Battery life
At least 15 hours at 50mA 300uS 50Hz
-
Adjustable Timer
10, 20, 30 45, 60, 90 min Defaults to 20 min
Fully shielded: touch proof mini USB
Output plug
Weight
90 gms without batteries
Dimensions
100 x 65 x 18 mm
Safety Classification
Internal power source. Designed for continuous use.
Type BF
Degree of protection against electrical shock.
IP22
No special moisture protection
Water resistance
Environmental Specifications
Humidity:
Operating:
Temperature range:
Atmospheric Pressure:
Storage:
Humidity:
Temperature range:
Atmospheric Pressure
The machine will often last for more than 5 years, the warranty
Expected service life
covers first 2 years. Accessories (leads, probes, pads, and
batteries) are not covered by the warranty.
•Lead life depends greatly on use. Always handle the leads
with care.
•Replace the probe every six months to ensure hygiene.
•Optional electrode pads should last 12-20 applications, depending
on skin condition and humidity.
•3.7V Li-ion Rechargeable Battery should last about 6 hours
continuous use.
• Storage life of an unopened pack of electrodes is 2 years.
Storage life
This may be affected by very high temperatures or very low
humidity.
• Storage life of batteries is 3 years.
• The unit and probe have no fixed shelf life.
37
elise au manual Rev 3.0 5/16 artwork.indd 37-38
.
20 to 93% RH
10 to 35C
700hPa to 1060hPa
Up to 93% RH non condensing
-20 to 70C
:
700hPa to 1060hPa
GMDN Definition : Stimulator, electrical, neuromuscular,
incontinence[36784]
A non-implantable neuromuscular electrical stimulator designed to
treat urinary and/or faecal incontinence that consists of a pair of
electrodes on a plug or pessary that are connected to a battery
powered pulse source. The plug or pessary is inserted into the
rectum or into the vagina and used to stimulate the muscles of the
pelvic floor.
IP22
The first number 2: Protected against access to hazardous parts
with a finger, and the jointed test finger of 12mmø, 80mm length,
shall have adequate
clearance from hazardous parts, and
protected against solid foreign objects of 12.5mmø and greater.
The second number 2: Protected against vertically falling water
drops when enclosure is tilted up to 15˚. Vertically falling drops
shall have no harmful effects when the enclosure is tilted at any
angle up to 15" on either side of the vertical.
PROGRAM SETTINGS
Program
Hz
Pulse
Ramp Up
Width
& Down
(µS)
(Sec)
STRES
50
300
1
URGE
10
200
1
MIXED
20
250
2
TONE
35
250
2
Default
Plateau
Off
Program
Duration
(Sec)
(Sec)
(Min)
5
10
20
5
10
20
5
20
20
3
6
20
38