NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head. Instruction Manual
M4×16 screws (pos. 2 in Fig. 2-2) with Allen key wo fix the panel on the
exchangeable mount.
Fig. 2-2. Replaceable panel mounted on the base unit
1 – screws for fastening to the base unit; 2 – screws for fastening to the exchangeable mount
This completes mounting the replaceable panel on the base unit.
2.2. Mounting the Optical Tube
To mount the optical tube on the base unit, perform the following steps:
1. If necessary, mount the scanner in the positioning device as follows:
a. Release the spring stops of the exchangeable mount (Fig. 2-3) and place the
scanner in the positioning device as shown in Fig. 2-4.
Fig. 2-3. Spring stops
b. Lower the scanner to its extreme by rotating the approach knob clockwise
(Fig. 2-5). This is necessary for mounting the optical tube.
18
Fig. 2-4. Scanner mounted in the positioning
device