NTEGRA Spectra Probe NanoLaboratory. SNOM Measuring Head. Instruction Manual
the measuring unit of the instrument (the base unit with the measuring head) should
be placed on a separate table away from computers and monitors, to eliminate
electromagnetic interference;
the table intended for installation of the measuring unit of the instrument must be
stable and, whenever possible, massive.
The operation of the device is susceptible to heat flows, draughts and sudden alternations
in temperature and humidity.

1.6. Storage and Transport Instructions

- Packaging of the equipment should be performed indoors in a ventilated room at
ambient temperature of 15° С – 40 °С and relative humidity at most 80 % with no
corrosive agent in the atmosphere. Storage conditions should comply with the
standard GOST 15150.
- The equipment should be carried by one or several modes of enclosed transport
according to transport regulations applicable to the mode(s) in service.
- Only transportation of the equipment in proper package is permitted to prevent
transportation damage.
- The equipment in its package should be placed and secured in a vehicle in a way
that provides steady position and excludes shocks of the deliverable parts by one
another or by walls of the vehicle.
- Transportation conditions should comply with requirements of the standards
GOST Р 52931-2008 and GOST 15150.
- Transportation and storage in regions of the Far North and equated localities
should comply with requirements of the standards GOST Р 52931-2008 and
GOST 15846.
- Location of the equipment in a warehouse must provide easy access to it and free
space for personnel.
- Devices and their parts should be stored on shelves.
- At least 100 mm distance between the stored parts and walls and ceiling of the
premises should be reserved.
- Distance between the stored parts and heating facilities of the premises should be
at least 0.5 m.