Trotec LD6000 Operating Manual

Combination detector
Table of Contents

Advertisement

Quick Links

LD6000
EN
OPERATING MANUAL
COMBINATION DETECTOR

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LD6000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Trotec LD6000

  • Page 1 LD6000 OPERATING MANUAL COMBINATION DETECTOR...
  • Page 2: Table Of Contents

    7. Acoustic leak and pipe detection LD6000 Measuring Device ......17 7.1. Parameter Settings in the Acoustic Modes ....7 LD6000 H2 hydrogen sensor .
  • Page 3: Safety Instructions

    We do not accept any liability for any damages which might occur as a result The LD6000 is a combo detector designed to allow the user to carry out elec- of improper use or the non-observance of these instructions. The guarantee...
  • Page 4: Control Elements And Connecting Ports

    LD6000 BM universal microphone 5.1. Microphone Jack and Sensor Port The LD6000 BM is a universal microphone which can be either be used as a First place the batteries into the battery compartment of the LD6000 and connect probe microphone or a contact microphone. Connect the LD6000 VL extension all the components which you require to perform your specific measuring operation.
  • Page 5: Connecting The Hydrogen Sensor For Trace Gas Detection

    LD6000. The LD6000 BM as a Probe Microphone: Connect the tip of the LD6000 VL to the base of the LD6000 BM either with or without the extension piece and connect the microphone to the microphone jack You are in the main menu when you power on the LD6000 for the first time.
  • Page 6: Time And Date

    With the LD6000 a frequency range of 0 - 4,000 Hz can be analyzed. Each of the measuring modes that can be used during acoustic leak detection has both a...
  • Page 7: Parameter Settings In The Acoustic Modes

    7.1.2. Automatic sensor sensitivity adjustment selected filter setting appears as the filter mode symbol In addition to the manual adjustment function, the LD6000 is also equipped with Press the cancel key to leave the window without changing the filter setting.
  • Page 8: Smart Mode

    The window for selecting the measuring mode will then open. To set the smart mode, navigate to the list item smart mode and confirm your selection by pressing the right-hand dial. The window will close and the smart mode symbol will be shown LD6000 Operating Manual...
  • Page 9: F&V Mode (Frequency And Volume)

    In addition, the sound value that is currently being measured is displayed measurement on Position 1 will be cleared so that the last measurement that in the form of a bar was taken last is always on Position 10. LD6000 Operating Manual...
  • Page 10: Pulse Mode

    , activate the symbol and confirm your selection. As many as 10 single measurements can be carried out with the LD6000 and used as a measurement series which can then be seen on the display. This series The measuring mode selection window opens.
  • Page 11: Adjusting The Filter And Switching On The Acoustic Function

    Save 7.7.3. Switching on the acoustic function during filter adjustment The LD6000 is equipped with a function which allows you to activate the acoustics while the filter is being adapted. 11 11 This function can be turned on or off by pressing the headphone key. The status...
  • Page 12: Parameter Settings For Acoustic Long-Term Measuring

    7.1.1. The LD6000 is excellently-suited for detecting leaks together with the optionally available LD6000 H2 hydrogen sensor and the formation gas type 95 /5, which Selecting the filter presetting comprises 95 % nitrogen and 5 % hydrogen.
  • Page 13: Display

    • Preset alarm value from, the LD6000 is also equipped with an acoustic feedback to help you during • Digital display with the current measuring value leak detection.
  • Page 14: Defining Alarm Thresholds

    LD6000 is equipped with a permanent alarm function with alarm thresholds ured value as reference value when zero calibration was carried out in the lower that can be configured individually.
  • Page 15: Saving And Loading Measuring Data

    Any measurement reading and any measurement series can be saved to the Measuring Data internal memory of the LD6000 permanently and either called up at a later point The measuring value you have selected is now displayed. or transferred to your PC.
  • Page 16: Troubleshooting

    • Replace the cable with a spare cable if possible or use the cable from a appears to be defective. second LD6000 to check to see if the sound receiver is working properly. 4. The cable that connects the • Check the connections between the connected headphones and the LD6000.
  • Page 17: Changing The Battery, Cleaning And Maintenance

    500 g Temperature conditions Storage: -25 °C to +65 °C The LD6000 H2 pump kit is a standard part of the LD6000 H2 hydrogen ground sen- Dimensions approx. L 210 x W 160 x H 60 mm sor (article no. 3.110.008.020). It can, however, also be ordered separately in Weight approx.
  • Page 18: Information For Use In The Field

    This method allows you to determine the section of pipe with the highest sound 25 Hz 250 Hz 2,500 Hz intensity without changing the control settings. The next leak detection stage can then be carried out on this particular section of pipe over ground. LD6000 Operating Manual...
  • Page 19: Schematic Leak Detection

    15.1.2.3 Pipe Detection with Pulse Wave Generator late the maximum amount of forming gas that is required to locate the leak: The LD6000 is equipped with a special PULSE mode (see chapter 7.5.) for acoustic pipe detection in combination with a separately available pulse wave generator, Formula: G = VL x L x D for example the LD-PULS.
  • Page 20 Trotec GmbH Grebbener Straße 7 D-52525 Heinsberg +49 2452 962-400 +49 2452 962-200 info @ trotec.com www.trotec.com...