Met One Instruments BX-807 Operation Manual

Pm2.5 sharp cut cyclone inlet

Advertisement

Quick Links

PM
Met One Instruments, Inc.
1600 NW Washington Blvd.
Grants Pass, OR 97526
Telephone: (541) 471-7111
Facsimile: (541) 471-7116
BX-807 Operation Manual - ©
publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any
form without the express written permission of Met One Instruments, Inc.
S
HARP
2.5
O
PERATION
www.metone.com
Copyright 2019 Met One Instruments, Inc. All Rights Reserved worldwide. No part of this
BX-807
C
C
UT
YCLONE
M
ANUAL
BX-807-9800 R
C
EV
I
NLET

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BX-807 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Met One Instruments BX-807

  • Page 1 BX-807 Operation Manual - © Copyright 2019 Met One Instruments, Inc. All Rights Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any form without the express written permission of Met One Instruments, Inc.
  • Page 2 Page 2 BX-807-9800 REV C...
  • Page 3: Table Of Contents

    Figure 2-1 Typical BX-807 Installation ....................6 Figure 3-1 BX-807 Air Flow Path ......................7 Figure 3-2 Exploded View of the BX-807 .................... 8 Table of Tables Table 3-1 Recommended Maintenance Schedule for the BX-807 ............7 BX-807-9800 REV C Page 3...
  • Page 4: Introduction

    Please have the instrument serial number available when contacting the manufacturer. On most models manufactured by Met One Instruments, it will be located on a silver product label on the unit, and also printed on the calibration certificate. The serial number will begin with a letter and be followed by a unique four or five digit number such as X12345.
  • Page 5: Sharp Cut Cyclone

    BX-807 cannot be considered as FRM 2.5 equivalent. For applications requiring this designation, the Met One Instruments, Inc. BX-808 or BX-809 size selective inlets must be used. The BX-807 may be used with any of the following Met One Instruments, Inc. products: • BAM 1020 • BAM 1022 •...
  • Page 6: Installation

    2.2 Instrument Assembly and Deployment The BX-807 is designed for easy deployment. It is installed on top of the sample inlet tube as a second stage after the BX-802 PM inlet. Follow the installation instructions of the instrument with which the SCC is being used.
  • Page 7: Operation And Maintenance

    PM10 separator, the sample air flows down into the top of the SCC, through the cyclonic separator, into the grit pot, and then exits the bottom of the BX-807 after passing through the top cap and transfer tube (see Figure 3-2 for component labels). The airflow path is shown in Figure 3-1.
  • Page 8: Disassembling The Bx-807

    3.2 Disassembling the BX-807 An exploded view of the BX-807 is shown in Figure 3-2. It demonstrates how the BX-807 comes apart for cleaning. Recommended cleaning materials are clean mineral free water and lint free lab wipes. In cases where stubborn deposits are observed, ultrasonic cleaning is recommended.
  • Page 9: Spare Parts

    6. Reinstall the SCC and then verify all O-rings are functioning properly. This should be done by performing a system leak check as described in the operation manual of the instrument being used with the BX-807. 4 SPARE PARTS The following parts are available from Met One Instruments, Inc. for maintenance. Consumables Description Part Number...