Please have the instrument serial number available when contacting the manufacturer. On most models manufactured by Met One Instruments, it will be located on a silver product label on the unit, and also printed on the calibration certificate. The serial number will begin with a letter and be followed by a unique four or five digit number such as X12345.
BX-807 cannot be considered as FRM 2.5 equivalent. For applications requiring this designation, the Met One Instruments, Inc. BX-808 or BX-809 size selective inlets must be used. The BX-807 may be used with any of the following Met One Instruments, Inc. products: • BAM 1020 • BAM 1022 •...
2.2 Instrument Assembly and Deployment The BX-807 is designed for easy deployment. It is installed on top of the sample inlet tube as a second stage after the BX-802 PM inlet. Follow the installation instructions of the instrument with which the SCC is being used.
PM10 separator, the sample air flows down into the top of the SCC, through the cyclonic separator, into the grit pot, and then exits the bottom of the BX-807 after passing through the top cap and transfer tube (see Figure 3-2 for component labels). The airflow path is shown in Figure 3-1.
3.2 Disassembling the BX-807 An exploded view of the BX-807 is shown in Figure 3-2. It demonstrates how the BX-807 comes apart for cleaning. Recommended cleaning materials are clean mineral free water and lint free lab wipes. In cases where stubborn deposits are observed, ultrasonic cleaning is recommended.
6. Reinstall the SCC and then verify all O-rings are functioning properly. This should be done by performing a system leak check as described in the operation manual of the instrument being used with the BX-807. 4 SPARE PARTS The following parts are available from Met One Instruments, Inc. for maintenance. Consumables Description Part Number...