Heat - K-Systems G85 Instruction Manual

Co2 bench top incubator
Hide thumbs Also See for G85:
Table of Contents

Advertisement

6.8 HEAT

The HEAT option is used for the heater to turn on at a certain time.
Note
This function only works in conjunction with St.St.
If St.St is set to 8.00 and heat function is on, the heater will
automatically turn on at 8.00 using 24 hour clock.
Follow these steps to navigate in the HEAT menu.
Action
Press and hold UP+DOWN key for
3 seconds to enter the menu.
The first menu "UNIT" will appear
in the display.
Press DOWN key once and the
menu "HEAT" appears in the
display.
Press and hold setpoint key to
change values.
While holding the setpoint key in,
press either arrow UP or arrow
DOWN key to select between
"ON" and "OFF".
When the required setting is
activated, let go of the setpoint
key.
Press and hold UP+DOWN key for
3 seconds to exit the menu.
Note
A dot will turn on in the right side of display, indicating that the Heat function
is active.
Key
23
Instruction Manual - G85
Display

Hide quick links: