Vibration - Greenheck Vektor-MH Installation, Operation And Maintenance Manual

Laboratory exhaust system
Table of Contents

Advertisement

Vibration

On start-up and during operation, the unit should
operate smoothly with minimal vibration. It is possible
that a higher degree of vibration may be experienced.
Excessive vibration if left unchecked, can cause a
multitude of problems, including structural and/or
1. Wheel Unbalance
2. Drive Pulley Misalignment
3. Incorrect Belt Tension
4. Bearing Misalignment
5. Mechanical Looseness
6. Faulty Belts
7. Drive Component Unbalance
8. Poor Inlet/Outlet Conditions
9. Foundation Stiffness
for corrective actions. If observation cannot locate the
source of vibration, a qualified technician using vibration
analysis equipment should be consulted. If the problem
is wheel unbalance, in-place balancing can be done
through the access panel located on the side of each
fan's tubular housing. Any correction weights added to
the wheel should be welded to either the wheel back
(single-plane balance) or to the wheel back and wheel
cone (two-plane balance).
®
component failure.
The most common
sources of vibration
are listed.
Many of these
conditions can be
discovered by careful
observation. Refer to
the troubleshooting
section of this manual
Laboratory Exhaust Systems
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vektor-mdVektor-ms

Table of Contents