OHAUS ST20M Series Instruction Manual

OHAUS ST20M Series Instruction Manual

Pen meter

Advertisement

Quick Links

Welcome to OHAUS!
After more than a century of perfecting the art of measurement through our durable weighing
products, OHAUS precision is now available in a line of portable electrochemistry products that
provides accurate measurement of pH, ORP, conductivity, TDS and salinity.
ST series pen meters are the economical option when you are simply looking for meters that are
easy to use and provide accurate measurements.
The pen meters should not be used by Children of age 12 or younger.
This product conforms to the RoHS Directive 2011/65/EU and the EMC Directive 2014/30/EU.
Thank you for choosing OHAUS waterproof pen meters. Please read the manual completely
before use. This manual serves the following models:
• ST20M-B (pen pH and Conductivity meter)
• ST20M-C (pen pH and Conductivity meter)
pH: Condition the pH electrode by immersing it in pH electrode storage solution (3M KCl
solution) for 1 hour before using. If the pH electrode protective cap is dry (may cause white
crystals), add storage solution in the cap and soak the pH electrode for 1 hour.
Conductivity: Rinse the conductivity electrode with pure water (distilled water) before using.
pH: ST20M has a 0.01 pH resolution and 0.05 pH accuracy and is factory calibrated.
You may need to calibrate the ST20M regularly to ensure the pH accuracy.
Conductivity: ST20M pen meter is factory calibrated, and requires calibration every week for
both pH and conductivity.
1.800.561.8187
ST20M Pen Meter
Instruction Manual
www.
.com
information@itm.com

Advertisement

loading

Summary of Contents for OHAUS ST20M Series

  • Page 1 Welcome to OHAUS! After more than a century of perfecting the art of measurement through our durable weighing products, OHAUS precision is now available in a line of portable electrochemistry products that provides accurate measurement of pH, ORP, conductivity, TDS and salinity.
  • Page 2 Remove the protective cap, rinse the pH electrode glass bulb with pure water (distilled water) and wipe clean. (Glass bulb is fragile, be careful not to scratch it.) Press the On/Off button to turn on the meter. Dip the electrode about 2 to 3 cm into the test solution (at least 20 ml). Stir several seconds, then stop and wait until the reading stabilizes.
  • Page 3 ST20M pen meter is factory calibrated, and requires calibration every week for both pH and conductivity. Be caution when using this function if you are not familiar with the conductivity pen meter. • Press the On/Off button to turn on the meter, and then press the Mode button to switch to conductivity mode.
  • Page 4 The electrode is replaceable. Model Content Item No. ST20M-B pen meter (electrode included), ST20M-B 30393199 pH buffer powder pH4.01 / pH7.00 / pH10.01 ST20M-C pen meter (electrode included), ST20M-C 30393200 pH buffer powder pH4.01 / pH7.00 / pH10.01 Description Item No. ST20M-B Electrode 30393203 ST20M-C Electrode...
  • Page 5 In lieu of a properly registered warranty, the warranty period shall begin on the date of shipment to the authorized dealer you purchased from. No other express or implied warranty is given by OHAUS Corporation. OHAUS Corporation shall not be liable for any consequential damages.