3 Analyzer Settings

Access the Wash Cycle Configuration menu – see Section 5 of the AX400 pH analyzer's User Guide (IM/AX4PH).
Cont.
-----
Pulsed
Wash Mode
Hours
15
Mins
Wash Frequency
Mins
15
Secs
Wash Duration
1. 0
Mins
Recovery Period
Continuous
Fig. 3.1 Wash Cycle Frequency
6
Wash Mode
Set to Cont. (continuous) for use with the dosing pump.
Wash Frequency
Set the wash frequency required.
Wash frequency is set in 15 minute increments between 15 and 45 minutes, then in 1 hour increments
between 1 and 24 hours.
Wash Duration
Set the wash duration required.
Wash duration is set in 15 second increments between 15 and 45 seconds, then in 1 minute increments
between 1 and 10 minutes.
Recovery Period
Set the recovery period required, between 0.5 and 5.0 minutes in 0.5 minute increments.
Wash Duration
Frequency
Recovery Period