Electrical Connections; Dosing Pump Mains Connections; Analyzer Connections; Dosing Pump Settings - ABB AAO100 User Manual

Dosing pump
Table of Contents

Advertisement

Dosing Pump
AAO100

2.2 Electrical Connections

2.2.1 Dosing Pump Mains Connections

The power supply range is 100 to 230 V AC.
Caution. Before making any electrical connections, ensure
that the power supply is switched off.
Referring to Fig. 2.2:
1. Connect the mains plug
2. Push the mains plug into the pump socket
3. Fasten the mains plug cover
A
Earth
Live
Neutral
Fig. 2.2 Dosing Pump Mains Connections
IM/DOSPMP Rev. B
A
.
B
.
C
.
B
C

2.2.2 Analyzer Connections

Referring to Fig. 2.3:
1. Plug the controller cable into the dosing pump
2. Cut the controller cable's Blue wire back to the outer
insulation.
3. Remove the analyzer's terminal cover plate – see Section
6.4.1. of the analyzer's User Guide (IM/AX4PH).
4. Twist the controller cable's Black and Brown wires
together and connect to terminal A10
5. Twist the controller cable's White and Grey wires together
and connect to terminal A12
6. Refit the analyzer's terminal cover plate – see Section
6.4.1. of the analyzer's User Guide (IM/AX4PH).
A
Fig. 2.3 Analyzer Connections

2.3 Dosing Pump Settings

For optimum cleaning, set the dosing pump to:
Stroke Length 100
Multifunction
External – the AX400 pH Analyzer controls
Switch
Refer to Section 3, page 6 for analyzer configuration
instructions.
2 Installation
A
.
B
.
C
.
AX400
Terminal block A
B
A9
Black
A10 C
Brown
A11 N/C
White
A12 N/O
Grey
A13
C
the frequency and duration of the
cleaning cycle.
3

Advertisement

Table of Contents
loading

Table of Contents