Network integration
94
Заказывайте на сайте: www.tinnova.ru/endress_hauser/ || Эл. почта: info@tinnova.ru
3
4
3
1
Transmitter with integrated WLAN antenna
2
Transmitter with external WLAN antenna
3
LED lit constantly: WLAN reception is enabled on measuring device
4
LED flashing: WLAN connection established between operating unit and measuring device
5
Computer with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for
accessing the integrated device Web server or with operating tool (e.g. FieldCare, DeviceCare)
6
Mobile handheld terminal with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft
Edge) for accessing the integrated device Web server or operating tool (e.g. FieldCare, DeviceCare)
7
Smart phone or tablet (e.g. Field Xpert SMT70)
Function
Encryption
Configurable WLAN channels
Degree of protection
Available antennas
Max. range
Materials:
External WLAN antenna
With the optional OPC-UA-Server application package, the device can be integrated into an Ethernet
network via the service interface (CDI-RJ45 and WLAN) and communicate with OPC-UA clients. If
the device is used in this way, IT security must be considered.
For permanent access to device data and for device configuration via the Web server, the device is
incorporated directly in a network via the service interface (CDI-RJ45). In this way, the device can be
1
4
2
5
6
WLAN: IEEE 802.11 b/g (2.4 GHz)
• Access point with DHCP server (default setting)
• Network
WPA2-PSK AES-128 (in accordance with IEEE 802.11i)
1 to 11
IP67
• Internal antenna
• External antenna (optional)
In the event of poor transmission/reception conditions at the place of
installation.
Available as an accessory → 104.
Only one antenna active in each case!
50 m (164 ft)
• Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-
plated brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Connector: Nickel-plated brass
• Angle bracket: Stainless steel
Proline Cubemass C 500
1
3
3
2
4
7
Endress+Hauser
4
A0034569
Need help?
Do you have a question about the Proline Cubemass C 500 and is the answer not in the manual?