9.7. PHOSPHATE HIGH RANGE

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93717A‑0
HI93717B‑0
REAGENT SETS
HI93717‑01
HI93717‑03
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the
Phosphate HR
procedure described in the Method Selection
section (see page 28).
• Follow procedures in Section 6 to prepare the
sample for analysis.
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for measurement.
60
0.0 to 30.0 mg/L (as PO
3‑
4
0.0 to 9.8 mg/L (as P)
0.1 mg/L
±1.0 mg/L ±4% of reading at 25 °C
LED with narrow band interference filter @ 525 nm
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Amino Acid method.
Description
Phosphate High Range Reagent A
Phosphate High Range Reagent B
Reagents for 100 tests
Reagents for 300 tests
method using the
)
Quantity
10 drops
1 packet
10 mL