Current Output Check - Ametek DR5200 Supplementary Instructions Manual

Radar fmcw level transmitter for liquids in storage and process applications
Hide thumbs Also See for DR5200:
Table of Contents

Advertisement

DR5200

8.5.2 Current output check

INFORMATION!
You can do part 2 of the proof test or do the procedures that follow. For more data about the
proof tests, refer to Proof tests on page 20
Do a check of the current output value: 4 mA
• Energize the device.
• Enter the Supervisor menu.
• Push [> > > > ], [ ] and [> > > > ] to go to menu item 2.2.1 SET OUTPUT. Push [> > > > ] and then push [ ] or [ ]
to move up or down the list to set the output to 4 mA.
i
The output will change to the set value after 5 seconds, as an alternative to the measured
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 4 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
The current output goes back to the measured value.
i
• End of the procedure.
Do a check of the current output value: 12 mA
• Push [> > > > ] and then push [ ] or [ ] to move up or down the list in menu item 2.2.1 SET OUTPUT
to set the output to 12 mA.
The output will change to the set value after 5 seconds, as an alternative to the measured
i
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 12 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
The current output goes back to the measured value.
i
• End of the procedure.
Do a check of the current output value: 20 mA
• Push [> > > > ] and then push [ ] or [ ] to move up or down the list in menu item 2.2.1 SET OUTPUT
to set the output to 20 mA.
The output will change to the set value after 5 seconds, as an alternative to the measured
i
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 20 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
i
The current output goes back to the measured value.
• End of the procedure.
CAUTION!
If the values do not agree with the tolerances, then the data supplied by device will not be
correct. For a solution to the problem, speak to the supplier.
DR5200-LM SIL2 Issue 2
.