Summary of Contents for Medivance Arctic Sun 5000

  • Page 1 O P E R A T O R ’ S M A N U A L...
  • Page 2: Table Of Contents

    O P E R A T O R ’ S M A N U A L Languages The 24/7 Helpline is intended to assist healthcare professionals English ..........................1 with technical questions they Český (Czech) ........................15 may have regarding the use of Danske (Danish) ......................29 the A ®...
  • Page 3: Chapter 1 - Getting Started

    Medivance recommends measuring patient temperature from a second site pressure reducing devices under the patient to protect from skin injury. to verify patient temperature. Medivance recommends the use of a second • Do not allow urine, antibacterial solutions or other agents to pool patient temperature probe connected to the A ®...
  • Page 4: System Setup

    3) A New User Training module option is available from the start up hypochlorite) as it may damage the system. screen. • Medivance will not be responsible for patient safety or equipment performance if the procedures to operate, maintain, modify or service the Medivance A ® Tem- RCTIC perature Management System are other than those specified by Medivance.
  • Page 5: Chapter 2 - Patient Therapy

    E N G L I S H Functional Verification Chapter 2 – Patient Therapy Perform the following functional verification procedure after initial setup and installation of the control module. Place A ™ Pads RCTIC 1) Power On the control module Read the Instructions for Use that accompany the A ™...
  • Page 6: Therapy Screens

    O P E R A T O R ’ S M A N U A L Therapy Screens Initiate Normothermia (Control Patient) Normothermia therapy is initiated and managed, and patient temperature Normothermia Therapy Screen is automatically controlled to a set target temperature from the Control Patient window in the Normothermia therapy screen.
  • Page 7: Initiate Hypothermia (Cool Patient And Rewarm Patient)

    E N G L I S H • To modify the patient target temperature and duration, press the Adjust button to display the Cool Patient-Adjust window. • Cool Patient To: Use the Up and Down arrows on the left side to set the desired patient target temperature to cool the patient Initiate Hypothermia (Cool Patient and Rewarm Patient) •...
  • Page 8: End Therapy

    O P E R A T O R ’ S M A N U A L • Using the Rewarm Patient From temperature, the Rewarm Patient To Chapter 3 – Normothermia Settings temperature and the rewarming rate settings, the system will calculate and display the rewarming duration and the date/time at which the the Normothermia Settings patient will reach the final rewarming target temperature.
  • Page 9: Chapter 4 - Hypothermia Settings

    E N G L I S H Chapter 4 – Hypothermia Settings Chapter 5 – Advanced Setup Hypothermia Settings Use the Advanced Setup screen to view the current settings and modify Use the Hypothermia Settings screen to view the current settings and the settings for the following parameters.
  • Page 10: Chapter 6 - Alarms And Alerts

    If either safety systems do not respond appropri- ately either an alarm 80 or 81 will be issued. Contact Medivance Customer External Surfaces Support.
  • Page 11: Software Update

    5000 Temperature Management System Service Manual for the full declaration regarding electromagnetic compatibility. Software Update Software updates will be provided from Medivance on a flash drive. Installation of software updates will be performed via the USB port on the Environmental Conditions front of the control module.
  • Page 12: Arctic Sun

    O P E R A T O R ’ S M A N U A L ® 5000 Temperature Management System Specifications RCTIC Parameter Specification Therapy Modes Normothermia: Control Patient Hypothermia: Cool Patient, Rewarm Patient Heater Capacity 2500 BTU/hr / 750 Watts Circulating Fluid Distilled or Sterile Water Reservoir Capacity...
  • Page 13: Appendix B: Symbols

    E N G L I S H Appendix B: Symbols The A ® Temperature Management System Control module bears the following symbols: RCTIC For the safe and effective use of this device, the operator must consult the accompanying documents prior to use Identifies European Representative This symbol adjacent to the patient connections means that the thermal probe connection is a “Defibrillator-Proof, Type BF Applied Part”, per standard IEC 60601-1 and affords the degree of...
  • Page 14: Appendix C: Electromagnetic Compatibility

    UT) for 5 cycles ® 5000 RCTIC magnetic environment specified below. The customer or the end user of the Arctic Sun 5000 IEC 61000-4-11 Temperature Manage- Temperature Management System should assure that it is used in such an environment. 70% UT (30% dip in...
  • Page 15 E N G L I S H 1.3 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 4: 1.4 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 6: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity Recommended separation distances between portable and mobile RF communications equipment and the A ®...