18. TECHNICAL SPECIFICATIONS

Output
Single channel
Max. 52 mA across 500 Ohm
Output current
Symmetrical Bi-Polar rectangular
Wave shape
14 mm x 24 mm LCD
Display
Battery
Rechargeable Li-ion 3.7V
Power supply
5V DC 1A
Dimensions
38x65.3x9.6 mm
Type BF
Degree of protection against electrical
Equipment
shock
IP22. No special moisture protection
Water resistance
At least 10 minutes
Contact Duration
Environmental Specifications
Operating:
Temperature range:
Humidity:
Temperature range:
Storage:
Humidity:
NB The electrical specifications are nominal and subject to
variation from the listed values due to normal production
tolerances.
Prog
Description
1
Constant Output
2
Constant Output
3
Burst Output 2Hz
4
Constant Output
33
5 to 40 C
15 to 93%RH non
condensing
-25 to +70 C
93%RH non condensing
Pulse duration
Rate Hz
μs
110
100
110
50
100
150
10
200

19. TROUBLESHOOTING

Problem
Possible
causes
No display Flat batteries
Low
Low batteries
battery
display
No
Incorrect
sensation
connection
Not strong
enough
Lead
faulty/damaged
Solution
Charge battery
Charge battery
Have you
applied both
electrode pads
to the Y-cable
to ensure a
complete
circuit?
Increase
strength. Most
users will feel
something at a
setting below
5.
Replace lead
34