General Specifications - TensCare itouch sure Instructions For Use Manual

Pelvic floor exerciser
Table of Contents

Advertisement

CHANGE OF LEAD
Please not that from serial number E14/009931 the iTouch Sure has
an all-plastic connecting plug on the lead to comply with current
safety regulations. For earlier models with a metal plug, use L-IT1
leads (please see images below).
19. DISPOSAL OF WASTE ELECTRICAL AND
ELECTRONIC PRODUCTS
At the end of its life, please treat this device as electronic waste and
dispose of responsibly according to current local regulations.
29
iTouch Sure User Manual UK Rev 3.2 9.16
20. GENERAL SPECIFICATION
50V zero to peak Constant voltage over 470-1500 Ohm
Max Intensity
Constant current over160-470Ohm
Channels
single
Waveform
Asymmetrical rectangular
Max Pulse energy *
Total output limited to25uC per pulse
Power
2 x AA Alkaline OR 2 x AA NiMH
Battery life
At least 15 hours at 50mA 300uS 50Hz
Adjustable Timer
- 10, 20, 30 45, 60, 90 min Defaults to 20 min
Pause button
"Clock" key when running
Output plug
Fully shielded: touch proof mini USB
Weight
90 gms without batteries
Dimensions
102 x 53 x 30mm
Safety Classification
Internal power source.
Type BF Designed for continuous use. IP22
Environmental Specifications:
Operating;
Temperature range:
10 to 35C
Humidity:
20 to 93% RH
Atmospheric Pressure:
700hPa to 1060hPa
Storage:
Temperature range:
-20 to +70C
Humidity:
93% RH noncondensing
Atmospheric Pressure:
700hPa to 1060hPa
Typical operation time:
No less than 1 week @ 50% AMP, 4 hours per day
TYPE BF
Equipment providing a degree of protection against
EQUIPMENT
electric shock, with isolated applied part.
This symbol on the unit means "Refer to Instructions for use"
The unit is not water resistant, and should be protected from liquids.
The machine will often last for more than 5 years, but is
Expected ServiceLife
guaranteed for 2 years. Accessories (leads, probes, pads, and
batteries) are notcovered by the warranty.
• Lead life depends greatly on use. Always handle the leads with care.
• Replace the probe every six months to ensurehygiene.
• Optional electrode pads should last 12-20 applications,
depending on skin condition and humidity.
• AA Alkaline Batteries should last about 18 hours continuous use.
Storage life
• Storage life of an unopened pack of electrodes is 2 years.
This may be affected by very high temperatures or very low humidity.
• Storage life ofbatteries is 3 years.
• The unit and probe have no fixed shelf life.
Made in China
30

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents