INSTRUCTIONS FOR USE TensCare Ltd, 9 Blenheim Road, Epsom, READ CAREFULLY BEFORE STARTING USE Surrey KT19 9BE, UK Email: info@tenscare.co.uk www.tenscare.co.uk iTouch Sure User Manual UK Rev 3.2 9.16 iTouch Sure User Manual UK Rev 3.2 9.16...
Leads Alarm Contents Installation of Batteries Using the “itouch Sure” Anal Probe Cleaning Battery Removal and Replacement Troubleshooting Consumables and Servicing Electrical Waste Disposal General Specifications General Specifications cont. Electromagnetic Interference Warranty iTouch Sure User Manual UK Rev 3.2 9.16...
• Usage Memory The “itouch Sure” has a Usage Memory which shows the length of time that the unit has been in use and the average level of strength utilised. iTouch Sure User Manual UK Rev 3.2 9.16...
2. TYPES OF INCONTINENCE 4. HOW CONTINENCE STIMULATION WORKS The “itouch Sure”, through a vaginal probe or pads, sends a gentle There are three types of incontinence: Stress, Urge, and Mixed. stimulation to your pelvis floor muscles using clinically recognised •...
Page 5
URGE INCONTINENCE: 5. CONTINENCE PROGRAMMES The “itouch Sure” has four preset programmes. One for each type of Shown on the screen as:“URGE” incontinence and a fourth for toning of the pelvic floor muscles. URGE incontinence programme reduces the involuntary contractions of the...
‘cramping’ feeling. If this happens, stop using your “itouch Sure” until the aching goes away, then start again using a lower strength and a shorter treatment time. As the pelvic floor begins to improve, it will be possible to maintain the stimulation forlonger.
When driving, operating machinery, or similar actions needing fine CAUTION: control. Loose pads, damaged leads, or sudden changes in contact • Observe caution when using the Itouch Sure at the same time as may cause brief involuntary muscle movements. being connected to monitoring equipment with body worn •...
Mix old, new or different types of batteries. • Use this device with leads, electrode pads, and accessories other than those recommended by the manufacturer in section 19. Strength Time Performance may vary from specification. DOWN Selector iTouch Sure User Manual UK Rev 3.2 9.16...
Page 9
Press and release to increase in soft and once. gentle 0.5 steps. The unit will remain in the WORK part of the cycle for 5 seconds while intensity is beingadjusted. iTouch Sure User Manual UK Rev 3.2 9.16...
Strength DOWNButton Time Selectorbutton To decrease strength: Each time you switch on the “itouch Sure” the length of time of your programme will Hold down button and strength will decrease automatically revert to 20 minutes. During use steadily in 0.5steps.
8. SCREEN INFORMATION 9. USAGE MEMORY The “itouch Sure” has a usage memory which remembers the length of time and average strength that has beenused. To activate hold down the Time Selector and strength UP buttons together for 3 seconds.
2 x 1.5v AA batteries • Storage pouch NB: Remove batteries from your itouch • SURE if the unit is unlikely to be used for Instruction booklet a long period. iTouch Sure User Manual UK Rev 3.2 9.16...
• Ensure the “itouch Sure” is switched OFF before insertion. • Lubricate the probe with a proprietary jelly, such as TensCare Go Gel or water based lubricant. • Insert the probe into the vagina, in the same way as a tampon, until only the flange at the end is visible.
You should consult your healthcare professional before starting treatment. 15. CLEANING The Liberty vaginal probe – that is supplied with the “itouch Sure” is Faecal Incontinence intended strictly for single person use. It is important that the probe is Faecal continence can be the result of weakened or poorly functioning cleaned after each use.
It ensures that rapid changes in connection cannot cause very uncomfortable rapid changes in stimulation. Unfortunately this means that some users, who fall outside of the general range, may experience unwanted LEAD alarms. iTouch Sure User Manual UK Rev 3.2 9.16...
If this happens to include you, you can try;- 18. CONSUMABLES AND SERVICING i) Using a water–based lubricant such as TensCare Go Gel, which will improve conduction. PART NUMBER ii) Crossing your legs and squeezing to increase pressure on the...
CHANGE OF LEAD 50V zero to peak Constant voltage over 470-1500 Ohm Max Intensity Please not that from serial number E14/009931 the iTouch Sure has Constant current over160-470Ohm an all-plastic connecting plug on the lead to comply with current Channels...
21. GENERAL SPECIFICATION (cont) The itouch Sure is intended for use in the electromagnetic environment specified below. The customer or the user of the itouch sure should assure that it is used in such an environment. PROGRAMME SETTINGS Emissions test...
Page 19
The Itouch Sure is intended for use in the electromagnetic environment specified below. equipment and the itouch Sure The customer or the user of the Itouch Sure should assure that it is used in such an environment. Immunity test IEC 60601 testlevel...