Sulfuryl Fluoride Monitor – GF1900
Doc 120-00007
Revision: G
Release Date: 11/04/2016
Page 1 of 20
901-0014 Rev A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GF1900 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Interscan Corporation GF1900

  • Page 1 GF1900 SULFURYL FLUORIDE MONITOR Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 1 of 20 901-0014 Rev A...
  • Page 2 This page intentionally left blank Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 2 of 20 901-0014 Rev A...
  • Page 3: Table Of Contents

    TABLE OF CONTENTS INTRODUCTION…………………………………………….....……..4 SYMBOL KEY ………………………………………………………………… 4 PICTURE OF GF1900 ………………………………………………………. 5 CASE LAYOUT…………………………………………………………..6-7 CAUTIONS…………………………………………………....…….… 8 OPERATION……………………………………………....…..…….…. 9 CALIBRATION………………………………………………………..... 10 MAINTENANCE………………………………………………....….….. 11 TROUBLE SHOOTING…………………………………………..…. 12-13 ACCESSORIES…………………………………………………..... 14-15 WARRANTY…………………………………………………..…....…. 16 DIAGRAMS, SCHEMATICS, AND PARTS......... 17-19 SCHEMATIC......………….......17 WIRING DIAGRAM ………………………....……..18 BATTERY PACK WIRING………….…....….….…….
  • Page 4: Introduction

    The furnace is a key and essential component of the GF1900. Due to extreme heat (900 degrees C) the furnace has a limited life. Mean Time Before Failure (MTBF) is one thousand (1000) hours.
  • Page 5 Figure 1 GF1900 instrument in blue carrying case (figure 1) Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 5 of 20 901-0014 Rev A...
  • Page 6 BAT TEST - analyzer is powered and meter displays battery pack level. Note: Meter level must be at or above LO BAT mark for proper operation. 11) Power Connector – input connector for Battery Pack or AC Power Supply. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G...
  • Page 7 (48) inches (121.92 cm) can be used only if calibrated with chosen length attached during process. NOTE: Longer lengths than twelve (12) inches (30.480 cm) can increase response time. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016...
  • Page 8: Cautions

    The presence of any of these could produce an EXPLOSION. 3) Areas which operate mobile phones, CB's, etc., may cause electronic interference. This is evidenced by random or erratic meter deflection. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016...
  • Page 9: Operation

    When the meter shows LO BAT, turn the control knob to OFF. 6) Recharge batteries overnight. Approximately nine (9) to ten (10) hours. 7) Approximately 70 minutes operating time is available with a fully charged battery Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016...
  • Page 10: Calibration

    This will generally cover the three samples. Readings can sometime jump when the sample bag is disconnected and then drop down. This is a normal condition which occurs due to changes in pressure. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G...
  • Page 11: Maintenance

    Note: Excessive force will result in furnace damage. Connect thermocouple wires and replace the sensor. Close the top to position the insulator so it does not lie against the furnace. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G...
  • Page 12: Trouble Shooting

    Thermocouple may be defective. Check across leads for mV reading. The reading should be greater than 20 mV after 5 minutes. Reverse thermocouple leads if the reading is negative. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016...
  • Page 13 FUNCTION SWITCH on, lift “Furnace Off” cover up and flip the switch up toward Meter. Check the meter for drift and adjust as necessary until stable. Position “Furnace Off” cover down when stability has been achieved. Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G...
  • Page 14: Accessories

    ACCESSORIES Figure 7 Figure 7: AC Power Supply - P/N 040-00079 (110VAC) , P/N 040-00080 (220VAC) Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 14 of 20 901-0014 Rev A...
  • Page 15 ACCESSORIES Con’t Figure 8 Figure 8: 24 VDC Battery Pack - P/N BP24V Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 15 of 20 901-0014 Rev A...
  • Page 16: Warranty

    Figure 9 Figure 9: Battery Charger - P/N 040-00077 WARRANTY INTERSCAN CORPORATION warrants the GF1900 gas fumigant analyzer of its manufacture (furnaces, sensors, batteries, fuses lamps, tubing, fittings, filters and scrubbers EXCEPTED) to be free from defects in material and workmanship for a period of one year from date of shipment.
  • Page 17 Additionally, in a custom system, warranty on any component shall not exceed the manufacturer's warranty given to INTERSCAN CORPORATION Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016...
  • Page 18: Diagrams, Schematics, And Parts

    DIAGRAMS, SCHEMATICS & PARTS Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 18 of 20 901-0014 Rev A...
  • Page 19: Wiring Diagram

    DIAGRAMS, SCHEMATICS & PARTS, Con’t. See insert Wiring Diagram Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 19 of 20 901-0014 Rev A...
  • Page 21 DIAGRAMS, SCHEMATICS & PARTS, Con’t. Parts 113-P Sulfuryl Fluoride (So2F2) Sensor Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 20 of 20 901-0014 Rev A...
  • Page 22: Return Authorization

    Below you will find the link to the online RMA request form: http://www.gasdetection.com/contact-interscan/rma-request/ For service information please contact: INTERSCAN CORPORATION Service Department (818) 882-2331 (800) 458-6153 (USA & CANADA) FAX: (818) 341-0642 Sulfuryl Fluoride Monitor – GF1900 Doc 120-00007 Revision: G Release Date: 11/04/2016 Page 21 of 20 901-0014 Rev A...