The AURA™ Sensor and Antenna Holster are type BF applied parts. The AURA™
Monitor, AURA
TM
are accessible parts.
Mishandling (such as dropping or crushing) could damage the AURA™ Monitor.
If you suspect damage to the AURA™ Monitor, contact your Branchpoint
Technologies representative or the customer service department for instructions
and return packaging.
The AURA™ Monitor has been classified with an International Protection Marking
of IP43. The AURA™ Monitor is therefore not waterproof or explosion-proof and
cannot be sterilized. Do not use it in the presence of flammable gas mixtures
containing anesthetics, oxygen, or nitrous oxide.
The AURA™ Monitor may be affected by electrostatic discharge (ESD). If ESD
occurs and the AURA™ Monitor's functionality is affected, attempt to reset the
AURA™ Monitor or contact Branchpoint Technologies for instructions.
AURA™ ICP Monitoring System has not been evaluated for pediatric use.
Avoid sources of ionizing radiation. If ionizing radiation must be used, avoid
focusing near the AURA™ Sensor placement site. Ionizing radiation may
damage the device, however, the damage may not be immediately detectable.
7. INSTRUCTIONS FOR USE
INSPECT THE PACKAGING
Visually inspect for mechanical and functional integrity of the packaging and its
contents as well as the legibility and adherence of the AURA™ Monitor labels. If
there is evidence of damage or tampering, do NOT use the product. Contact
Customer Service, and Branchpoint Technologies will provide instructions for
return.
Package contents include:
AURA™ Monitor (1)
AURA™ Antenna (1)
Lithium Ion 11V rechargeable batteries (2)
Battery Charger (1)
Power Supply (1)

CONTROL PANEL

4
Antenna, batteries, battery charger, and AURA
Figure 2: AURA™ Monitor Controls.
Data Receiver