Safety
WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS SUCH AS
ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER THAN 30 CM (12
INCHES) TO ANY PART OF THE EMS 20, INCLUDING CABLES SPECIFIED BY THE MANUFACTURER.
OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.
WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED OR
PROVIDED BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT IN INCREASED ELECTROMAG-
NETIC EMISSIONS OR DECREASED ELECTROMAGNETIC IMMUNITY OF THIS EQUIPMENT AND RESULT IN
IMPROPER OPERATION.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER EQUIPMENT SHOULD
BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION. IF SUCH USE IS NECESSARY, THIS
EQUIPMENT AND THE OTHER EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERATING
NORMALLY.
Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect
Medical Electrical Equipment.
Caution: The EMS 20 should not be used adjacent to or stacked with other equipment and that
if adjacent or stacked use is necessary, the EMS 20 should be observed to verify normal opera-
tion in the configuration in which it will be used.
Table 1
Guidance and Manufacturer's declaration—electromagnetic emissions
The EMS 20 is intended for use in the electromagnetic environment specified below. The customer or the
user of the EMS 20 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment—guidance
Group 1
The EMS 20 uses RF energy only for internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The EMS 20 is suitable for use in all establishments, including
Not applicable domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
Not applicable buildings used for domestic purposes.