ROUTe:MONitor (@<ch_list>)
ROUTe:MONitor?
Description: Select the monitor channel. (@<ch_list>) is channel list containing only
one channel number. When monitoring is enabled and scanning is active, the monitor
channel will be measured and displayed continuously during the time between scan
sweeps.
This command does not initiate scanning.
Example:
Response:
Set Example: ROUT:MON (@101)
Related Commands:
INITiate
ROUTe:MONitor:DATA?
ROUTe:MONitor:STATe
ROUTe:MONitor:DATA?
Description: Query the latest reading of the monitor channel. At least one measurement
must have been acquired from the monitor channel. If no monitor data is available, an
error is logged (603,"Data not available"). *RST disables monitor and clears monitor
data.
Example:
Response:
Related Commands:
ROUTe:MONitor
ROUTe:MONitor:STATe
INITiate
ROUTe:MONitor:STATe <boolean>
ROUTe:MONitor:STATe?
Description: Enable or disable channel monitoring during scanning. <boolean> can be 0
(off), OFF, 1 (on), or ON. When monitoring is enabled and scanning is initiated, a
designated channel will be measured and displayed continuously during the time between
scan sweeps. *RST disables monitor.
Example:
Response:
Set Example: ROUT:MON:STAT ON
Related Commands:
ROUTe:MONitor
www.elso.sk
ROUT:MON?
101
ROUT:MON:DATA?
1.000000e+00
ROUT:MON:STAT?
1