Fluke HYDRA III Series Remote Programmers Manual page 58

Data acquisition unit
Hide thumbs Also See for HYDRA III Series:
Table of Contents

Advertisement

2638A
Remote Programmers Guide
MMEMory:LOG:READ[:DATA]? ["<name>"]
Description: Read a data file stored in USB memory. <name> is the name of a stored
data file folder. If the named file does not exist or there are no data files and an error is
logged (901, "File not found"). If the <name> parameter is omitted, the latest data file is
read.
If USB memory is not available,an error is logged (1603, "USB memory not ready"). The
response contains the entire contents of the data file arranged in rows of data separated by
semicolons. Each row consists of a series of data elements separated by commas. A data
element may be a number or a text string. Text strings are enclosed in quotation marks.
*RST does not affect USB memory.
Example:
Response:
Related Commands:
DATA:LOG:ENABle?
DATA:LOG:DESTination?
MMEMory:LOG:STORe
MMEMory:LOG:CATalog?
MMEMory:LOG:READ:CONF? [<name>]
Definition: Read the configuration of a data file stored in USB memory. <name> is the
name of a stored configuration file. If the named file does not exist or there are no data
files an error is logged (901,"File not found"). If the <name> parameter is omitted, the
latest data file is read. The response contains the entire contents of the configuration file.
*RST does not affect USB memory.
Example:
Response:
Related Commands:
MMEMory:LOG:READ[:DATA]?
MMEMory:LOG:STORe
Description: Store the latest scan data to USB memory. If no scan data is available, no
action is taken and an error is logged 603,"Data not available"). If USB memory is not
available or is write protected and an error is logged (1603,"USB memory not ready").
If the amount of data exceeds the maximum number of rows, the data is stored in
multiple associated files. The name of the file created using this command can be queried
using the MMEMory:LOG:CATalog? command. *RST does not affect USB memory.
Example:
Related Commands:
DATA:LOG:ENABle?
DATA:LOG:RLIMit
MMEMory:LOG:READ?
MMEMory:MOUNt?
54
www.elso.sk
MMEM:LOG:READ?
1.000000e+00,1.000000e-01,
MMEM:LOG:READ:CONF?
The entire contents of the configuration file.
MMEM:LOG:STOR