Remote Programming Examples
This section gives examples of sequences of commands for several likely scenarios.
Single Channel, One-Shot
In this scenario, it is desired to measure one channel on command in a one-shot fashion.
This example measures the temperature of a type K thermocouple using internal
reference junction compensation. Below are two methods that can be used. In the first
method, the channel is configured and read in one operation. The second method
configures the channel and reads the channel in separate operations.
Command
MEAS:TEMP? TC,K,(@102)
MEAS:TEMP? TC,K,(@102)
Command
CONF:TEMP TC,K,(@102)
READ?
READ?
www.elso.sk
Configure channel 102 to measure the temperature of a type K
thermocouple using internal RJC. Enable channel 102 only
then set the trigger source to timer and the trigger count to 1.
After configured, acquire the measurement and return the
result.
Acquire another temperature measurement and return the
result.
Configure channel 102 to measure the temperature of a type K
thermocouple using internal RJC. Enable channel 102 only
then set the trigger source to timer and the trigger count to 1.
Acquire one measurement then return the reading.
Acquire another measurement then return the reading.
Hydra Series III Data Acquisition Unit
Remote Programming Examples