Write protection

4.4
The transmitter parameters can be protected against accidental or unauthorised overwriting. The write
protection prevents changing the parameters using the keys or PC/laptop or HARTR Communicator. But
they can be read out by PC/Laptop or HARTR Communicator.
-
Clean the transmitter before opening it to prevent the ingress of dirt
-
Switch off the voltage for transmitters conforming to protection type "Flame--proof enclosure" when
using in hazardous location, undo the screw holding the safety angle on the electronics side (if fitted)
-
Move angle out of the way
-
Screw off housing cover of the electronics side
-
On the LCD (optional) is the displayed letter
-
Secure jumper against losing: push jumper horizonta
-
Screw on housing cover
-
Fix safety angle (if applicable) and switch on the voltage
*)