About this document Memocheck Sim CYP03D About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
Repairs not described in the Operating Instructions provided must be carried out only directly at the manufacturer' s site or by the service organization. Designated use Memocheck Sim CYP03D is a test tool for analysis measuring points. It enables the simulation of user-definable measured values and errors of all sensors incorporating Memosenstechnology.
• ATEX design approval BVS 12 ATEX E 008 with amendments The EC Declaration of Conformity is an integral part of these Operating Instructions. The Memocheck Sim CYP03D is powered by three alkaline batteries with a total nominal voltage of 4.5 V.
Page 7
IECEx Connecting the approved digital Memocheck Sim CYP03D sensor simulator to the IECEx- certified, intrinsically safe sensor output circuit of the Liquiline M CM42 (IECEx TUR 11.0007X) transmitter or alternatively to an IECEx-certified, intrinsically safe Memosens...
Page 8
Basic safety instructions Memocheck Sim CYP03D ‣ The transmitter' s Memosens input must support the following maximum values. In particular, the effective inner inductance and the capacitance of the approved, intrinsically safe sensor output may not exceed these values: 1. Entity Parameters 2.
Device description Memocheck Sim CYP03D Simulation values With the Memocheck Sim CYP03D you can simulate the following data: • Simulation values – Main values – Raw values – Temperature • Parameter – pH glass (pH glass) – pH glass, SIL sensor (pH glass SIL) –...
Memocheck Sim CYP03D Incoming acceptance and product identification Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved.
Dieselstraße 24 D-70839 Gerlingen 4.2.5 Scope of delivery Memocheck Sim CYP03D • Memocheck Sim CYP03D • Operating Instructions • 1 quality certificate as ordered • Cable as per order (optional) • Case to store CYP03D and cable (optional) • Certificate of calibration (optional) Certificates and approvals 4.3.1...
Memocheck Sim CYP03D Electrical connection Electrical connection Connection in hazardous areas only with ATEX- Approval Ex ia CYK10 CYP03D not permitted A0026198-EN 2 Operation in hazardous areas Connecting the simulator The electrical connection of the sensor' s to the transmitter is established using measuring cable CYK10.
Operation options Memocheck Sim CYP03D Operation options Overview 6.1.1 Display Menu path and/or device designation Battery status Simulated parameter Assignment of soft keys, e.g. menu Assignment of soft keys, e.g. and Sim. main value A0026099-EN 4 Display (e.g. simulation mode) 6.1.2...
Page 15
Memocheck Sim CYP03D Operation options and Reset Press MINUS and ENTER key simultaneously for longer Saved setups remain intact. and Factory settings Press PLUS and ENTER key simultaneously for longer • This resets the device to the factory settings.
Operation options Memocheck Sim CYP03D Structure and function of the operating menu 6.2.1 Menu structure Language Display Illumination Contrast Setup Simulation Parameter Memosens pH glass pH glass SIL pH Isfet pH + ORP Cond c Cond c 4-pol Oxygen (amp.) Oxygen (opt.Memo.)
Page 17
Memocheck Sim CYP03D Operation options 6.2.2 Operating concept Example of changing values within a menu: defining the start value of a ramp Maximum and minimum values are displayed in the editor. You can only configure values within these limits. Use the arrow keys to select the digit of the value that you want to change.
Commissioning Memocheck Sim CYP03D Commissioning Switching on the measuring device Switch on the device ‣ Press and hold for at least 3 seconds. This loads the last setup that was saved. Switch off the device ‣ Press and hold for at least 3 seconds.
Page 19
Memocheck Sim CYP03D Commissioning You can now simulate the selected parameter using the selected settings. Once you connect the Memocheck Sim CYP03D to a Memosens transmitter, the Memosens icon appears in the status line of the simulator. It indicates that the simulator is communicating with the transmitter.
Operation Memocheck Sim CYP03D Operation Configuring the measuring device 8.1.1 Display behavior Possible settings • Illumination • Contrast Path: Menu/Display ‣ Change the illumination or contrast setting for the display using or . Confirm entry with .
Page 21
Memocheck Sim CYP03D Operation 8.1.3 Selecting parameters In the setup, you configure which parameters are to be simulated. Path: Menu/Setup/Simulation/Parameter. The parameters currently configured are displayed. Select new parameter: Parameter/Memosens or Fixed cable. Select the desired parameter. If the parameter is changed, the following prompt appears: ESC = cancel ...
Page 22
Operation Memocheck Sim CYP03D Menu/Setup/Simulation/Test values: set other measured values, which are shown on the display ( Temperatureonly) or on the transmitter as a fixed value. Change the value to the desired display, e. g. 25.00 °C. The test value remains at the value set here and cannot be modified in the simulation mode.
Memocheck Sim CYP03D Operation Sim main values Memosens (factory settings in bold) pH glass pH glass SIL pH Isfet pH + ORP • pH value • pH value • pH value • pH value • Temperature • Temperature • Temperature •...
Page 24
If you change a calibration value, there will be a brief interruption to communication between the Memocheck Sim CYP03D and the transmitter to allow the transmitter to accept the calibration settings. An unfavorable configuration may result in measured values outside the specified measuring range.
Page 25
Memocheck Sim CYP03D Operation y2-y1 y2-y1 x2-x1 x2-x1 A0017356 A0017359 5 Linear function 6 Negative slope Zero point Zero point Slope Slope Start Value Start Value Stop value Stop value Time Time x2-x1 = duration x2-x1 = duration 8.2.4...
Page 26
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Glass SCS failure Once you have selected an error using , Temp. sens. defect the error is output at the transmitter. To Glass SCS warning delete the error again, remove the check mark again using .
Page 27
Memocheck Sim CYP03D Operation Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Leak. curr. alarm Once you have selected an error using , Leak. curr. warn the error is output at the transmitter. To Temp. sens. defect...
Page 28
Operation Memocheck Sim CYP03D 8.2.6 Extended functions: pH + ORP Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value pH value -200.0 mV pH comp. isoth. 0.0 to 12.0 pH 7.0 pH mV comp. isoth. -300.0 to 300.0 mV 0.0 mV...
Page 29
Memocheck Sim CYP03D Operation Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Glass SCS failure Once you have selected an error using Ref. SCS failure , the error is output at the Temp. sens. defect transmitter.
Page 30
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Temperature Temperature -40.0 to -40.0 to 10 to 6000 s 1 to 200 0.5 to 600 s 150.0 ˚C 150.0 ˚C 60 s -40.0 °C 150.0 °C Raw value temp.
Page 31
Memocheck Sim CYP03D Operation Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Temp. sens. defect Once you have selected an error using , Sensor supply the error is output at the transmitter. To delete the error again, remove the check mark again using .
Page 32
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Cond c Once you have selected an error using , • Polarizat. warn. the error is output at the transmitter. To • No Cond. display delete the error again, remove the check •...
Page 33
Memocheck Sim CYP03D Operation 8.2.9 Extended functions: Oxygen (amp.) Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Slope 0.1 pA/hPa to 5.0 nA/hPa 313.5 pA/hPa Zero point -3.2 nA to 3.2 nA 0.0 pA Temp. offset -10.0 to 10.0 ˚C...
Page 34
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Leak. curr. alarm Once you have selected an error using , Leak. curr. warn the error is output at the transmitter. To Temp. sens. defect...
Page 35
Memocheck Sim CYP03D Operation 8.2.10 Extended functions: Oxygen (opt.Memo.) Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C Temp. gradient -3.0 to 3.0 Cal. Quality 0.0 to 100.0 % 100.0 %...
Page 36
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Measurement stop Once you have selected an error using , No ref. cal. P1 the error is output at the transmitter. To No ref. cal. P2...
Page 37
Memocheck Sim CYP03D Operation 8.2.11 Extended functions: Oxy. (opt.fixed) Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Slope 0 ... 200 % 100% -5.0 to 105.0 µs 20.0 µs Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C Temp.
Page 38
Operation Memocheck Sim CYP03D Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Error simulation Tau too low Once you have selected an error using , Tau too high the error is output at the transmitter. To No signal dropout...
Page 39
Memocheck Sim CYP03D Operation 8.2.12 Extended functions: Chlorine Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Slope -500.0 to -0.01 nA/g/l -25.0 nA/g/l Zero point -100.0 nA to 100.0 nA 0.0 pA Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C...
Page 40
Operation Memocheck Sim CYP03D 8.2.13 Extended functions: Turbidity Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C Error simulation LED error Once you have selected an error using , Turbidit.
Page 41
Memocheck Sim CYP03D Operation 8.2.14 Extended functions: Nitrate Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C Error simulation Flash lamp defect Once you have selected an error using , Turbidit.
Page 42
Operation Memocheck Sim CYP03D 8.2.15 Extended functions: SAC Path: Menu/Advanc. function Function Configuration options (factory settings in bold) Calibration value Temp. offset -10.0 to 10.0 ˚C 0.0 ˚C Error simulation Flash lamp defect Once you have selected an error using , Turbidit.
Only ever open the battery compartment in the non-hazardous area! 10.3 Calibration and qualification The Memocheck Sim CYP03D can, with the quality or calibration certificate, also be used as a qualification tool for your measuring point. The quality and calibration certificates can be renewed: For such services, you will need to return Memocheck Sim CYP03D to Endress+Hauser.
The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
71128721 CYK10-G032 + adapter; only for CYP03D, Ex To connect Memocheck Sim CYP03D to transmitters with M12 sockets and Pg couplings, you require the Memosens data cable CYK10 supplied. The cable is always supplied with an adapter piece so that it fits both M12 sockets and Pg couplings. If you wish to simulate fixed cable sensors (turbidity, nitrate, toroidal conductivity, oxygen optical) with Memocheck Sim CYP03D, you need this cable.
Technical data Memocheck Sim CYP03D Technical data 13.1 Environment 13.1.1 Ambient temperature range –20 to 50 °C (–4 to 120 °F) 13.1.2 Storage temperature -20 to 55 °C (-4 to 130 °F) 13.1.3 Degree of protection IP55 13.1.4 Electromagnetic compatibility...
Memocheck Sim CYP03D Technical data 13.2 Mechanical construction 13.2.1 Dimensions 94.0 (3.7) 40.0 (1.57) 25.0 (0.98) 58.0 (2.28) A0026005 7 Memocheck Sim CYP03D All dimensions in mm (inch) 13.2.2 Weight (incl. batteries) 0.3 kg (0.7 lbs) 13.2.3 Materials Housing:...
Page 48
Memocheck Sim CYP03D 13.2.4 Batteries For the Memocheck Sim CYP03D, use only the following battery types, as only these are covered by the Ex approval: • Energizer, EN91 (AA, 1.5 V, LR6 as per IEC), x 3 • Battery storage temperature: -20 to 35 ˚C (-4 to 95 ˚F)