Power Supply / Terminal Diagram - Endress+Hauser Memograph M RSG40 Technical Information

Graphic data manager. record, visualize, analyze and communicate
Hide thumbs Also See for Memograph M RSG40:
Table of Contents

Advertisement

Electrical connection (wiring
diagram)
R
L
To modem: Cable with
14
25 pol. Sub-D socket
1
2
3
To modem: Cable with
6
9 pol. Sub-D socket
1
To PC: Cable with
2
3
25 pol. Sub-D
14
plug
1
2
3
To PC: Cable with
6
9 pol. Sub-D plug
1
RxD - 3
GND - 5
2
3
TxD - 2
8

Power supply / terminal diagram

8
9
Further units
Power supply
115 / 230 VAC
24 V AC/DC
50 / 60 Hz
50 / 60 Hz
L+ N- PE
Rel. max. 250 V / 3 A
RS 232
25
7
13
5
8 9
9
5
5
8 9
25
7
13
5
8 9
9
Ethernet
5
LED
LED
5
8 9
green
yellow
RTD
I
+
U
-
a) > 1 V
TC
Pulse/
frequency
R =
Cable
L
resistance
RxD/TxD(+)
RxD/TxD(-)
Analog
outputs (O)
(Selectable in unit setup)
1
2
Pulse
+ - + -
1
2
0...20 mA;
+ - + -
4...20 mA
>20 ms
5
1
9
6
USB
RS 232 / RS 485
Bus Interface
x1
x2
x3 x4 x5 x6
x1
x1
+
-
x1
x1
x2
x3 x4 x5 x6
+
-
b)
1 V
x1
+
-
x1
+
>30 µs
-
RS 485
Digital in (D)
12V-24VDC
>5 mA
_
+
24 V -
GNDx
GNDx
Dx1
10k
24 V +
-
+
24V
24V Out: max. 200 mA
Out
-
Power supply slot
Slot 5
analog input 17-20 or digital-I/O (option)
Slot 4
analog input 13-16 (option)
Slot 3
analog input 9-12 (option)
Slot 2
analog input 5-8 (option)
Slot 1
analog input
1-4 (option
Interface slot
(option)
Analog inputs
x2
x3 x4 x5 x6
x1
x2
x3 x4 x5 x6
x2
x3 x4 x5 x6
0...20 mA, 4...20 mA, 0...5 mA
(Volt drop
1 V, R i
50 Ohm)
x2
x3 x4 x5 x6
a) 0...5 V, 0...10 V, +/- 10 V, +/- 30 V
b) 0...1 V, +/- 150 mV, +/- 1 V
x2
x3 x4 x5 x6
J, K, T, N, L, D, C, B, S, R
x2
x3 x4 x5 x6
LOW = 0...7 mA
HIGH = 13...20 mA
Memograph M
_
Dx1
12V-24VDC
+
(standard)
)
Pt50, Pt100,
Pt500, Pt1000,
Cu50, Cu100
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents