Section 8: Service; Software Accessible Service; Alarm Log; Fault Indicators - FALEX 400 Operation And Maintenance Manual

Fuel thermal oxidation test machine
Table of Contents

Advertisement

8.1 Software Accessible Service

Obtaining manual access to machines functionalities is
done through the 'Service' screen accessed from the
maintenance screen
[Figure 70]
various controls so operation should only be undertaken
by someone with understanding of the system.
1. I/O & Vars:
[Figure 71]
controls can be accessed by pressing the large
green or red buttons under Digital Points. The
HPLC pump can manually be set to flow at either 3
or 9 ml/min for this service screen. Live feedback is
shown in the Data section of this screen.
2. Close DP Neg: Pressing this button manually
operates the DP negative valve. This button
is not interlocked with the bypass valve and
should be used only if the operator has an
understanding of the system as not to damage the
pressure transducer.
Service Screen Reached From
Figure 70
The Maintenance Screen
I/O & Variable Screen
Figure 71
. The user has access to
Manual operation of different
3. Close DP Pos: Pressing this button manually
operates the DP positive valve. This button is not
interlocked with the bypass valve and should be
used only if the operator has an understanding
of the system as not to damage the pressure
transducer.
4. Factory: This is a protected area for the authorized
users only.

8.2 Alarm Log

Pressing the alarm log button displays the last
[Figure 72]
25 warnings and abort alarms.

8.3 Fault Indicators

On the front of the Falex 400 are 17 indicator lights. If
a fuse trips an indicator light will be visible with the
description of the problem area. Cycle the power on the
unit and restart the procedure that caused the fault. The
fuse should have reset. If the indicator light comes on
again, contact the Falex service department.
Listing of most current alarms
Figure 72

Section 8: Service