FALEX 400 Operation And Maintenance Manual page 36

Fuel thermal oxidation test machine
Table of Contents

Advertisement

Section 6: Maintenance
35
4. Grasp the desiccant cartridge with one hand
and pull the desiccant cartridge straight out of
the machine. If rubber bushings are attached
to desiccant filter, remove and replace in
machine
[Figure 55]
desiccant cartridge.
5. Obtain a new desiccant cartridge
6. Align the new desiccant cartridge with the
desiccant cartridge inlet/outlet ports. Push
the desiccant cartridge straight into the F400
desiccant housing until the cartridge snaps into
the retainer clips.
7. Power on the machine, and proceed to
maintenance screen. Push the aeration button.
Verify the humidity level is dropping. If it is not,
the desiccant is not seated correctly in the rubber
bushing. Remove desiccant cartridge, reposition
rubber bushing and reinstall cartridge.
8. Once the cartridge is properly installed. Replace
desiccant cover and install six screws to secure.
Desiccant Cartridge Housing
Figure 55
. Properly dispose of used
.
[Figure 56]
Desiccant Cartridge
Figure 56
Desiccant Housing
Retainer Clips
Desiccant Cartridge Inlet/Outlet Ports
Rubber Bushings