FALEX 400 Operation And Maintenance Manual page 38

Fuel thermal oxidation test machine
Table of Contents

Advertisement

37
6.3.2 Priming Pump
6.3.2.1 Manual Method
Although not the recommended method, the pump and
lines may be primed manually.
1. Open the sliding door in front of the pump by
gently pressing the black handle in and sliding the
door to the left
2. Place a towel under the two pump heads.
3. On the main screen, select the MAINTENANCE
button.
.
[Figure 60]
Section 6: Maintenance
4. On the Maintenance screen, select PRIME PUMP
button.
5. Press the PRIME button.
6. Insert an empty, plastic 10 ml syringe into the
pump priming port at the center of the black
priming port knob
[Figure 61]
port knob counterclockwise and suction fuel out
of the pump priming port by pulling in the syringe
plunger. Allow 6–8 ml fuel to fill syringe. Close
the black priming port knob by turning clockwise.
Remove the syringe from the black priming port
knob and dispose of fuel appropriately.
Black Handle
Opening the Door
Figure 60
to Prime the Pump
Black Priming Port Knob
Plastic Syringe
Two Pump Heads
Priming the Pump
Figure 61
. Rotate black priming

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for FALEX 400

Table of Contents