Maxtec MaxO2ME Operating Manual & Instructions For Use page 5

Hide thumbs Also See for MaxO2ME:
Table of Contents

Advertisement

NOTE: The
MaxO
ME
2
adjustable down to 15% that requires deliberate action to set it below 18%.
See section 3.1 Alarm Setting Procedure.
DO NOT clean or dry the MaxO
the MaxO
ME may damage components and render the system inoperable.
2
DO NOT over clean the MaxO
on critical components. Excessive residue buildup can affect the MaxO
» When cleaning the MaxO
MaxO
ME in liquid sterilizing agents or liquids of any kind.
2
directly onto the device.
DO NOT sterilize the MaxO
» If the MaxO
ME does not function as outlined in section 2.0, contact a Maxtec trained service
2
technician or Maxtec for service.
DO NOT allow the sensor to come in contact with exhaled patient gases or other potential sources of
contamination. The sensor face cannot be decontaminated if it comes in contact with infectious agents.
» Gas leaks that cause room air to mix with the gas sample may cause inaccurate oxygen readings.
Ensure the O-rings on the sensor and flow diverter are in place and intact prior to use.
DO NOT expose the sensor face to liquids or allow humidity to condense on the face of the
sensor as this may impair the function of the MaxO
» The MaxO
ME and sensor are non-sterile devices.
2
» Regularly inspect the MaxO
DO NOT use if damaged.
DO NOT obstruct alarm.
DO NOT smoke in an area where oxygen is being administered.
» The MaxO
ME may only be calibrated using 20.9% oxygen (room air) or 100% oxygen. Calibration at
2
other concentrations will result in inaccurate readings.
» The MaxO
ME may be used for ground transport but should not be used during air transport as
2
changes in barometric pressure will alter the monitor readings.
» The oxygen sensor should be operated in an upright position (sensor face downwards).
Operating the oxygen sensor upside down may cause the sensor to function improperly.
» When using the approved external power supply, functional batteries must also be installed in the
device. The device will not operate solely on the external power supply.
oxygen monitor has been manufactured with a low alarm setting
ME with a high pressure air gun. Applying high pressure air to
2
ME. Repeated use of a cleaning agent can cause residue buildup
2
DO NOT use harsh abrasives.
ME:
2
DO NOT allow cleaning solution to pool on the device.
ME. Standard sterilization techniques may damage the monitor.
2
ME and associated components for damage or electrolyte leakage prior to use.
2
866.4.Maxtec www.maxtec.com
ME's performance.
2
DO NOT immerse the
DO NOT spray cleaning solution
ME.
2
III