brewer Access 5000 Manual

Exam table
Hide thumbs Also See for Access 5000:

Advertisement

Quick Links

Important Information ............................................2
Safety Information .................................................3
Component Identification ......................................3
Specifications ........................................................4
Electrical Requirements ........................................4
Document #100578 RevG
Model 5000/5001
Installation .............................................................5
Operation ..............................................................7
Maintenance ........................................................13
Limited Warranty .................................................16
1
Printed in USA © 2010

Advertisement

Table of Contents
loading

Summary of Contents for brewer Access 5000

  • Page 1: Table Of Contents

    Model 5000/5001 Important Information ..........2 Installation .............5 Safety Information ..........3 Operation ..............7 Component Identification ........3 Maintenance ............13 Specifications ............4 Options, Accessories & Common Service Parts .15 Electrical Requirements ........4 Limited Warranty ..........16 Document #100578 RevG Printed in USA © 2010...
  • Page 2: Important Information

    Brewer Model 5000/5001 examination during general examinations conducted by qualified table, call the Brewer Customer Service Department at medical professionals. Read and understand all safety 1.888.Brewer.1. Our trained staff will attempt to assist information, operating instructions, and maintenance you in correcting the problem directly over the phone.
  • Page 3: Safety Information

    SAFETY INFORMATION Safety Information Failure to follow instructions with this symbol The primary concern of The Brewer Company is that the could result in serious personal injury and/or damage equipment is operated and maintained with the safety of to the unit.
  • Page 4: Specifications

    SPECIFICATIONS Weight of Table .............220 lbs. Electrical Requirements (With carton and skid)...........265 lbs... . .120 VAC nominal, 60 Hz, 3 amps maximum ... . including maximum 2 amp accessory outlet Table Weight Capacity ..........500 lbs.
  • Page 5: Installation

    INSTALLATION Overview 1. Carefully remove outer carton by removing the sta- ples along the bottom of the carton and lift off table. Perform the following sequence in order when setting up the table: NOTE: Save all packaging in case re-shipment is re- quired.
  • Page 6 INSTALLATION (CONTINUED) Leveling the Table (Figure 4) A leveling screw pad is located at each corner under the table’s base. Adjust the four leveling pads, by turning them up or down, to achieve a solid, level installation. Installing the Paper Rod (Figure 5) A dowel rod for supporting a paper roll is shipped with each table.
  • Page 7: Operation

    OPERATION General Back Section (Figure 6) 1. Press and hold the release handle, lift back section upward to raise. Releasing the handle will automati- DANGER cally lock the table in the desired position. Do not use this table in an explosive or oxygen- rich atmosphere.
  • Page 8 OPERATION (CONTINUED) Footstep (Figure 7) 1. Place hand into opening in front of step and gently lift while pulling step out to desired position. WARNING 2. To retract step to stowed position, gently lift while pushing step fully back into cabinet. Extend the footstep far enough to allow enough surface area for patients to safely stand on the WARNING...
  • Page 9 OPERATION (CONTINUED) Stirrups (Figure 8) Retracting the stirrup to the storage position: 1. Rotate stirrup to the first lateral position (parallel to WARNING the table). Failure to engage the stirrup lateral locking mecha- 2. Fold stirrup down against bar. nism could allow patient to lose balance resulting in personal injury to patient.
  • Page 10 OPERATION (CONTINUED) Leg Extension (Figure 9) WARNING Horizontal Positioning The leg extension is not intended to be extended during entry onto or exit off of the table. The leg extension is not intended to be used as a seat. Failure to heed this warning could result in serious personal injury.
  • Page 11 OPERATION (CONTINUED) Patient Positioning (Figure 10) The table is designed to accommodate the following ex- amination positions: • Full Flat Table Position • Chair Position • Lithotomy (Pelvic) Position Chair Position Lithotomy (Pelvic) Position Full Flat Table Position Figure 10. Patient Positioning Document #100578 RevG Printed in USA ©...
  • Page 12 OPERATION (CONTINUED) Pelvic Lift - Model 5001 Only (Figure 11) Drawer Warmer - Model 5001 Only Raising the seat section to the Pelvic Lift Position: (Figure 11) CAUTION 1. Raise the front edge of the seat section. The drawer warmer may not be touched when 2.
  • Page 13: Maintenance

    MAINTENANCE Preventative Maintenance Painted Metal Surfaces Wipe all painted metal surfaces with a clean cloth at least • Periodically inspect the electrical cord to ensure it is once a week. free of cuts or damage. Drawers • Inspect the mechanical functions to ensure satisfac- Regular care should be maintained by wiping out drawers tory operation.
  • Page 14 MAINTENANCE (CONTINUED) Fuse Replacement (Figure 12) WARNING Replace fuse with same type and rating only (P/N 100091, 3A, 250V). 1. Push bottom rear drawer to the left to access fuse assembly. See Figure 12. 2. Push fuse in, turn counterclockwise and pull outward. See Figure 13.
  • Page 15 OPTIONS, ACCESSORIES, AND COMMON SERVICE PARTS Listed below are the accessories which are authorized for use with this table. Description Order No. Use/Restrictions Replacement Upholstery 5115-XX Intended Use - Replacement top with hardware for 5000 and 5001, examination tables in 18 standard colors.
  • Page 16: Options, Accessories & Common Service Parts .15 Limited Warranty

    (a) the design, manufacture, sale, delivery, installation, repair, operation or use of products of The Brewer Company or any part thereof, (b) any actual or alleged failure or defect in products of The Brewer Company or any part thereof, or (c ) any actual or alleged breach or non-performance by The Brewer Company of this limited warranty.

This manual is also suitable for:

Access 5001

Table of Contents