User Supplied Equipment Requirements - Labcyte Echo 650 Series Quick Start Manual

Liquid handler
Table of Contents

Advertisement

SITE PREP GUIDE | Echo® 650 Series Liquid Handler
Vacuum source for source
plate dryer (2 options)
Note: A standalone Vacuubrand MZ 2S-NT vacuum system can be purchased from Labcyte (Part #001-8104). For more
information, see the
Note: The 10 L surge tank comes standard with the Echo Liquid Handler to ensure consistent vacuum.
Instrument mounting surface
(minimal requirements)
Destination plate samples
Coupling Fluid

User Supplied Equipment Requirements

The user must provide the following items:
Hardware
Client PC
Note: Client PC can be purchased directly from Labcyte (Part #ECHO-Client Computer). For information on the Client PC
available from Labcyte, see the
Option 1: House supply. Minimum 200 Torr (266 mbar, 22 in Hg Vac) minimum pumping speed
3
8.0 m
/hr
Option 2: Standalone vacuum with surge tank (10 L). 100 Torr (133 mbar, 26 in Hg Vac),
minimum pumping speed 2.0 m
Vacuum Pump (Optional and sold separately by Labcyte)
The Echo Liquid Handler can emit vibration while in operation and is susceptible to vibration
from other devices. For this reason, the Echo system should be placed on a surface that is
isolated from surrounding equipment or designed to suppress the transmission of vibration
from devices. The surface must be sturdy, level, and water-resistant with cross-bracing or
welded joints.
The surface must be capable of supporting a minimum of 142 kg (313 lbs) weight with
sufficient surface area to accommodate the Echo Liquid Handler in the recommended
operating envelope (see Echo Liquid Handler Dimensions). If the mounting surface is a table
top, the table legs should not rest on casters. If the table has casters, they should be locked.
If the chiller will be installed on the bench surface, it must accommodate an additional 14 kg
(31 lbs).
At installation, ensure common destination plates to be used with the instrument are
available for tuning the destination plate gripper.
Use only ASTM D-1193 Type 1 water in the fluidics system
Coupling Fluid Temperature: 22.1°C ± 0.9°C (71.8°F± 1.6°F)
Compressed air fitting compatible with 0.170 in ID / 0.250 in OD tubing between the compressed
air supply and the Echo Liquid Handler tubing (Labcyte provides the compressed air line and
fitting with the Echo Liquid Handler)
Vacuum fitting that connects 3/8 in ID, 5/8 in OD tubing to either house vacuum or a vacuum
pump. The supplied vacuum line is HercoBraid, 3/8 in ID, 5/8 in OD (RyanHerco P/N 0512-072)
Minimum Requirements: Microsoft® Windows® 7 using 32- or 64-bit OS, or Microsoft® Windows®
10 64-bit; Intel® Core™ 2 Duo processor or later; 1GB of RAM or greater; 1 GB of available disk
storage; display resolution of 1280 x 1024 (1920x1020 preferred); 2 x Gigabit ethernet ports
Client PC (Sold separately by Labcyte)
3
/hr
topic.
topic.
PN | 001-16480
3