Page 4
700M HFID 6-2-15 Communication and Interface ..................... 144 Analog and Digital Interface .................... 145 Serial Interface........................ 149 AK Protocol ........................150 Instruction Command ......................152 Error Handling ........................154 General AK Requirements ..................... 155 Scan Commands ........................157 Control Commands ........................ 168 Configuration Commands ......................
6-2-15 Introduction Thank you for purchasing the CAI 700M HFID Analyzer. Before using the 700M HFID, please familiarize yourself with its operation by reading this manual. If you have any questions, please do not hesitate to call California Analytical Instruments Technical Support for assistance. We want you to be among our thousands of satisfied customers.
Page 6
RS232 or TCP/IP interface, or automatically when the analyzer is placed in the Auto Range. The 700M HFID is available with an optional internal heated sample pump, and internal zero and span solenoids. It includes a backlit 3 x 5 inch liquid-crystal display and a 20-key data/operation...
Page 7
6-2-15 Operating Principle The CAI 700M HFID methane/non-methane hydrocarbon analyzer uses the principle of Flame Ionization Detection (FID) to determine the hydrocarbons within a gaseous sample. For methane analysis, a solenoid valve diverts the sample through a cutter assembly that removes all the hydrocarbons from the sample except methane.
Page 8
700M HFID 6-2-15 Flow System Combustion air and fuel used by the instrument are controlled by an Electronic Proportional Control (EPC) valve whose function is to maintain a constant pressure for combustion air at the inlet to a capillary. The pressure is factory-set for optimum analyzer performance.
Effect: Less than 2% Effect: Less than 1.2 to propane Flow Control: Electronic proportional pressure controller Sample Flow Rate: Typically 2.0 to 2.5 LPM (consult CAI for other flow rates) Fuel Requirements: 40% H /60% He (250cc/min). (Approx. flows) Fuel Inlet Pressure: 25 psig...
Page 10
700M HFID 6-2-15 Oven Temperature: 191°C Ambient Temperature: 5 to 40°C Ambient Humidity: Less than 90% RH (non-condensing) Warm-up Time: 1 hour (typical) Fittings: ¼-inch tube Power Requirements: 115 VAC/60 Hz or 230 VAC/50 Hz ± 10%, 750 Watts max.
700M HFID 6-2-15 Installation Safety Information Safety Alert Temperature Hazard Electrical Shock Hazard Caution or Warning Caution or Warning Caution or Warning Note, Caution and Warning symbols appear on the instrument and in this manual to draw your attention to important operational and safety information.
Page 12
700M HFID 6-2-15 symbol (a lightning bolt in a triangle) precedes an electrical shock hazard CAUTION or WARNING statement. Some or all of the above symbols may appear in this manual or on the equipment. This manual should be consulted whenever one of these symbols is encountered on the equipment.
Page 13
6-2-15 Fuel Requirements The CAI factory configures the 700M HFID analyzer for either 100% Hydrogen or 40%/60% Hydrogen/Helium fuel. Please make sure to use the correct fuel as specified on the fuel label affixed to the back panel of the analyzer.
Page 14
700M HFID 6-2-15 2. Always use a second wrench on the body of the filter housing when attempting to inspect or replace the filter. Never attempt to disassemble or reassemble the filter housing while it is hot. Always allow the filter housing to cool to room temperature before attempting any maintenance.
Page 15
IMPORTANT: You should save the original shipping container your analyzer arrives in. The shipping container and packaging are specially designed to protect the analyzer in transport. If you ever need to return the analyzer to CAI for repair or any other reason, the original shipping container and packaging should be used.
Page 16
700M HFID 6-2-15 Rack Mounting The front panel is designed for mounting into a standard 19-inch rack enclosure. Holes are located on the left and right side to allow the panel to be secured in the rack by screws. Optional rack slides allow the analyzer to be pulled out of the rack enclosure for access.
Page 17
5. The instrument is designed for rack mounting. Optional rack-mount slides are available. 6. Do not install the 700M HFID near equipment that emits electromagnetic interference (EMI). NOTE: A front and rear supporting brace or equivalent are required if the optional rack mount slides were not purchased.
Page 18
700M HFID 6-2-15 Electrical All wiring is connected at the rear of the analyzer. The AC power cord is connected to the power entry as shown below: Fuse Panel Power ON/OFF Switch Power Entry AC Power Switch, Connector and Fuse.
Page 19
700M HFID 6-2-15 Calibration gases can be introduced through either the calibration ports on the back of the analyzer (if optional solenoid valves have been installed) or through the sample inlet. Gases introduced through a calibration port should be at 20-25 PSIG. If introduced through the sample port, pressures should be as follows: Without sample pump, pressure should be 10-25 PSIG.
Page 20
, HCl etc. are present in the sampled gas. Gas Temperature The 700M HFID temperature is factory set at 191ºC unless specified otherwise by the customer. When measuring high-temperature gas streams, do not exceed the instrument’s maximum temperature rating of 410ºF (210ºC).
Page 21
700M HFID 6-2-15 If the analyzer does not contain the optional internal sample pump, the sample gas entering the instrument should be between 8 and 25 PSIG with a minimum flow capacity of 3 liters/min. If the analyzer does contain the optional sample pump, DO NOT pressurize sample inlet.
Turn on the Power switch on the analyzer’s rear panel. After the analyzer is turned on, it needs at least 30 seconds for initialization. After a short delay, the digital display should illuminate. If the display does not come on, check the power source and the fuse. If the problem persists, call CAI Technical Support.
Page 23
700M HFID 6-2-15 Proper Storage After power down, allow the heated analyzer components to cool to room temperature before preparing for storage. If the original shipping box was retained, the analyzer should be stored in the box in the packing material supplied.
700M HFID 6-2-15 Using the Keypad When the Measure screen is displayed, the ten Function keys (F1 through F10) are shortcuts to commonly used screens. On other screens, these keys can either be used as function keys or to enter numeric values. This is why each number key includes both the larger Function number at the top (for example, F1) and the smaller number underneath for numeric value (for example, The light key is used to turn the display’s backlight on and off.
Page 25
700M HFID 6-2-15 From any screen, the Meas key takes you to the Measure screen. The current measurement is being displayed. The Enter key: 1. In Function mode, the Enter key selects the highlighted function. 2. When a field is highlighted for numeric input, pressing the Enter key opens the selected field for numeric entry with a blinking cursor.
Menu Flow Chart The menu flow chart is a handy reference that will help you familiarize yourself with the operation of the CAI System 700M HFID Analyzer. Start by pressing to access the Main Menu to quickly find any screen.
Page 27
700M HFID 6-2-15 Calibration Menu Main Menu Menus Calibration Manual Automated Cal Gas Calibration Calibration Reset Factory Factors Calibration Calibration Concentration Setup Settings Initiate Manual Auto Cal Reset Zero Sequenced Cal Deviations Schedule Offset and Gain Zero Gas Initiate Span...
Page 28
700M HFID 6-2-15 Range Setup Menu Main Menu Menus Range Setup AutoRange AutoRange Range Limits On/Off Switch Points AutoRange On AutoRange Off Security Level Legend Security Level 1 STANDARD Security Level 2 SETUP Security Level 3 FACTORY...
Page 29
700M HFID 6-2-15 Diagnostics Menu Main Menu Menus Diagnostics Diagnostic Raw Value I/O Status Status Line Values Display Analyzer Digital Turn Status Line ← Temps → Outputs ← Press → ← → ← Flows → Analyzer Digital Turn Status Line...
Page 30
700M HFID 6-2-15 Setup Menu Main Menu Menus Setup Data Logging Auto Start Measure Output TCP/IP Clock Time Settings Settings Settings Parameters Averaging Programmable Programmable Time Analogs Digitals Output Output Output Output THC/CH Assignments Assignments Test Hold/Clear NMHC Mode Times...
Page 31
700M HFID 6-2-15 Alarms Menu Main Menu Menus Alarms Alarm Display Current Alarms Alarm Log Alarm Limits On/Off Turn Alarm Temperatures Display On Turn Alarm Pressures Display Off EPC Voltages Concentrations Security Level Legend Security Level 1 STANDARD Security Level 2...
Page 32
700M HFID 6-2-15 Service Menu Main Menu Menus Service Linear Factory Service Coefficients CAI use Only Range 1 Range 2 Range 3 Range 4 Security Level Legend Security Level 1 STANDARD Security Level 2 SETUP Security Level 3 FACTORY...
Page 33
700M HFID 6-2-15 Security Menu Main Menu Menus Security Operator Levels Reset Password Change Password Standard Functions Setup Functions Factory Functions Security Level Legend Security Level 1 STANDARD Security Level 2 SETUP Security Level 3 FACTORY...
The Main Menu is your gateway to operational, setup and maintenance functions on the 700M HFID analyzer via the corresponding function keys. All software functions of the 700M HFID analyzer can be reached via the menu above from the Main Menu screen. Operation starts by pressing the Menu key to bring up the Main Menu.
700M HFID 6-2-15 Measure Screen Screen Name Current Range AK Status Line Measured Concentration Second Log On Analyzer Mode Shortcut Menu Range Limit Current Time & Active Alarms Keypad Status The Measure Screen provides a visual of the current concentration of the gas being analyzed, along with other pertinent information.
Page 36
700M HFID 6-2-15 Screen Name: The name of the active screen the Analyzer is in; in this case the Measure screen. Second Log On: SEC appears when the second log is enabled. See Data Logging Time. AK Status Line: When the AK Status line is enabled, it will scroll through the analyzer’s present state using AK Protocol.
Page 37
700M HFID 6-2-15 Diags Diagnostic Values is used to view Temperatures, Pressures, EPC Percent Full Scale and Flows. Auto Range Allows operators to turn Auto Range On or Off. Manual Cal The Manual Calibration menu allows operators to Zero or Span the analyzer.
700M HFID 6-2-15 Measure Mode → → The Measure Mode menu is used to select one of three measurement modes: THC, CH THC/CH /NMHC. The Measure Mode menu is accessed by pressing the key on the Main Menu. This menu will affect how the analyzer operates and what is displayed in the Measure screen.
700M HFID 6-2-15 THC Mode → → To move to the THC mode, press while in the Measure Mode menu. In THC mode, the sample gas does not pass through the analyzer’s non-methane cutter. The final reading is Total Hydrocarbons.
700M HFID 6-2-15 THC/CH /NMHC Mode → → The THC/CH /NMHC ode activates the “sample and hold” feature which allows the analyzer to automatically cycle between THC and CH measurement. To change to the THC/CH /NMHC mode, press from the Measure Mode menu. The current measuring mode (cycle) is indicated above the analyzer range.
6-2-15 Analyzer Info → The Analyzer Info screen contains the basic identity of your 700M HFID Analyzer. The Analyzer Info screen is accessed by pressing the key on the Main Menu. This screen includes the Model and Serial Number of your analyzer (for easy identification if you are discussing your analyzer with CAI), factory settings for Sample pressure, Fuel pressure and Air pressure, and the software versions being used.
700M HFID 6-2-15 Remote/Manual → The Remote/Manual menu gives the operator the ability to control the instrument manually using the keypad or via a remote computer. The Remote/Manual menu is accessed by pressing the key on the Main Menu. The current setting (Remote Mode) is displayed in the upper right-hand corner of the screen.
→ When the analyzer is in Standby Mode, the pump is turned off and the solenoid valves are closed. The CAI logo is displayed along with the Serial Number. Standby mode is accessed by pressing the key from the Main Menu.
700M HFID 6-2-15 Ignition → Before Ignition is attempted the oven temperature should be a minimum of 120° C. To start the ignition sequence, press from the Main Menu. At the start of the ignition sequence the fuel valve will open five seconds before the air valve to prime the burner.
700M HFID 6-2-15 Menus → → The Menus screen provides access to most instrument features, including Calibration, Setup and Diagnostics. From the Main Menu press to bring up the Menus screen. Press to access the Calibration menus. Press to access the Range Setup menu.
6-2-15 Calibration → → The 700M HFID Analyzer requires initial calibration with zero and span calibration standards before operation. These calibrations can be performed manually or initiated automatically. From the Menus screen press to access the Calibration menu. The Calibration menu includes important features including basic setup for both manual and automated calibration.
700M HFID 6-2-15 Manual Calibration Whether you are calibrating a single range or multiple ranges, each range requires its own complete zero and span calibration. If you are calibrating multiple ranges during one session, the zero calibrations can all be performed before any of the span calibrations, as long as they are within the same relatively short time period.
700M HFID 6-2-15 Manual Calibration THC or CH Mode → → → → When a manual calibration is performed in Single Mode operation (THC or CH mode) the analyzer will remain in that mode during calibration. In Single Mode operation the analyzer can only be calibrated for one mode.
700M HFID 6-2-15 Zero → → → → Zero calibration should be performed before a span calibration. From the Manual Calibration menu press to access the Manual Zero Calibration screen. Make sure the analyzer is in the range you wish to calibrate. Use the Up/Down arrows to go to the desired range.
Page 50
700M HFID 6-2-15 When the concentration value has stabilized, press to set the zero calibration. The zero portion of the calibration should now be complete. If the calibration was successful, the screen will say ***Saved Current*** above the concentration value.
Page 51
700M HFID 6-2-15 If the zero calibration is unsuccessful, check the following: 1. Make sure the correct gas was introduced into the analyzer. 2. Verify the Diagnostic Values while flowing gas. 3. Check Zero Gas Deviations under Manual Deviations: →...
700M HFID 6-2-15 Span → → → → A span calibration should be performed after a successful zero calibration. From the Manual Calibration menu press to access the Manual Span Calibration screen. Make sure the highlighted span gas value (see above) matches the value on the certificate for the span calibration gas being supplied to the analyzer.
Page 53
700M HFID 6-2-15 When the concentration number has stabilized, press to set the span calibration. The calibration should now be complete. If the span calibration was successful, the screen will say ***Saved Current***. If the calibration was unsuccessful, the screen will say Outside Deviation Limits. See the Manual Zero Calibration section for examples of screens showing successful and unsuccessful calibrations.
700M HFID 6-2-15 Manual Calibration THC/CH /NMHC Mode → → → → When performing a calibration in Switching mode (THC/CH /NMHC mode), the operator has the option of using basic calibration in THC mode or advanced calibration to create Correction Factors using THC and CH modes.
700M HFID 6-2-15 Automated Calibration → → → An automated calibration is a timed zero calibration immediately followed by a timed span calibration. The Automated Calibration menu offers two choices: Sequenced Calibration and Sequenced Check of the existing calibration. The Automated Calibration menu is accessed by pressing from the Calibration menu.
Page 56
700M HFID 6-2-15 • This automated calibration is triggered manually and not by the analyzer’s clock or via remote signal. A fully automated sequenced calibration can be preset to include the desired interval for recurring analyzer-initiated calibrations. This requires additional setup.
700M HFID 6-2-15 Initiate Sequenced Cal → → → → Current Concentration Analyzer Mode Expected Span Concentration Countdown Timer Current Date and Time Current Status Because of timing requirements, sequenced calibrations are generally only used when the analyzer is controlling the flow of zero and span gases into the analyzer. To initiate a sequenced calibration, press from the Automated Calibration menu.
Page 58
700M HFID 6-2-15 1. Zero Range 1 Purging – Allows time for the zero gas to flush out any residual gases that may still be present in the detection path. 2. Zero Range 1 Calibrating – The calculated averaged zero is set as the new offset value,...
700M HFID 6-2-15 Initiate Sequenced Check → → → → Initiate Sequenced Check is a useful tool for setting up Auto Calibration. From the Auto Calibration Menu screen, pressing initiates a sequenced calibration check. Rather than initiating a calibration, it checks the validity of your most current calibration.
700M HFID 6-2-15 Cal Gas Concentrations → → → → The Cal Gas Concentrations Screen allows operators to change calibration gas values for multiple ranges on one screen. To access the Cal Gas Concentrations screen (shown above) press from the Calibration Menu.
700M HFID 6-2-15 Calibration Setup → → → The Calibration Setup menu provides all the parameters necessary for completing a successful calibration. To access the Calibration Setup menu, select from the Calibration menu. All parameters on the Calibration Setup menu apply to Automated Calibration. The following also apply to Manual Calibration: Cal Pump/Valves, Auto Cal/Check, Deviation Limits and Analog Hold on Cal.
700M HFID 6-2-15 Auto Calibration Schedule → → → → Next Scheduled Scheduled Screen Name Calibration Time and Date Calibration Interval Save Hourly, Daily or Current Date and AutoCal On / Off Weekly Time The Auto Calibration Schedule screen allows the operator to run automated calibrations using the analyzer’s internal clock.
Page 63
700M HFID 6-2-15 After all the changes have been made, you must choose one of the following: (Hourly), (Daily) or (Weekly) to save your changes. If this is not done, the selected changes will not be made and the analyzer will revert to the previous settings.
Then select a range (from 1 – 4) for calibration. To select all ranges, press 0. Press to close the field. The second parameter the operator can change is the Mode. The 700M HFID analyzer can calibrate in either THC or CH mode. Press...
Page 65
700M HFID 6-2-15 Calibration Gases gives you a choice of calibrating with Zero gas only or All calibration gases (zero and span gases). Press to open the field and change the parameter using the up or down arrows . Press to close the field.
700M HFID 6-2-15 Calibration Via Pump/Valves → → → → The use of Cal via Pump/Valves depends upon how calibration gases are being introduced into the analyzer – via a sample pump or internal valves (if equipped with internal valve option).
700M HFID 6-2-15 Auto Calibration/Check → → → → Auto Calibration/Check lets the operator select whether the analyzer actually calibrates, or performs a check of the calibration. To access the Auto Cal/Check menu, press from the Calibration Setup menu. The current setting is shown on the upper right corner of the screen.
700M HFID 6-2-15 Auto Calibration Timing → → → → Auto Calibration Timing determines the length of time it takes the analyzer to perform the Zero and Span cycles during a sequenced auto calibration. To access the AutoCal Timing screen, press from the Calibration Setup menu.
Page 69
700M HFID 6-2-15 2. Calibrating Time: during this 10-second time, the analyzer will calculate new offset and gain factors. The calibrating time is factory-set at 10 seconds and cannot be changed by the operator. 3. Verifying Time: during this time the measured value is checked to make sure it does not...
700M HFID 6-2-15 Deviation Limits → → → → Deviation Limits are used by the operator to define the maximum acceptable error limits of the zero and span gases for both manual and sequenced calibration. To access the Deviation Limits menu, press from the Calibration Setup menu.
700M HFID 6-2-15 Maximum Calibration Error → → → → → Maximum Calibration Error is used by the operator to define the maximum acceptable tolerances for Absolute and Relative deviations. Each range has its own set of Absolute and Relative tolerances. The deviations must be inside these tolerances to accept a calibration. To access the Maximum Calibration Error screen, press from the Deviation Limits menu.
700M HFID 6-2-15 Maximum Verifying Error → → → → → Maximum Verifying Error is the allowable tolerance during the Verifying step of sequenced calibration. To access the Maximum Verifying Error screen press from the Deviation Limits menu. To set the allowable tolerances for different ranges, use the up or down arrow move the highlight to the field you intend to change.
700M HFID 6-2-15 Analog Hold on Cal → → → → Analog Hold on Cal will hold the analog outputs to the last measured value during calibration. If Analog Hold on Cal is Off the analog outputs will be live. The existing setting (Off) is shown at the top right of the menu.
700M HFID 6-2-15 Calibration Factors → → → Calibration Factors allow the operator to track and view changes from the factory and previous calibrations. To access the Calibration Factors menu, press from the Calibration menu. Press to view the Manual Calibration Deviations menu.
700M HFID 6-2-15 Manual Deviations → → → → The Manual Deviations menu allows the operator to view the Zero and Span Deviations from Manual Calibrations. Press from the Calibration Factors menu to access the Manual Deviations menu. Press to view Zero Gas deviations.
700M HFID 6-2-15 Zero Gas Deviations → → → → → Press from the Manual Deviations menu to view the Zero Gas Deviations screen. Absolute Zero Gas Deviation is the zero gas content calculated by the factory polynomial related to the calibrated range limit.
700M HFID 6-2-15 Span Gas Deviations → → → → → Press from the Manual Deviations menu to view the Span Gas Deviations screen. Absolute Span Gas Deviation is span gas bottle value minus span gas value calculated by the factory polynomial related to the calibrated range limit.
700M HFID 6-2-15 Auto Cal Deviations → → → → The Auto Calibration Deviations menu gives the operator a choice of viewing either zero or span verifying deviations. The verifying deviations are taken during the verifying stage of sequenced and auto calibrations. Press from the Calibration Factors menu to access the Auto Cal Deviations menu.
700M HFID 6-2-15 Verifying Zero Deviations → → → → → Press from the Auto Cal Deviations menu to view the Verifying Zero Deviations screen. Measured Value is the averaged concentration during the Verifying Zero stage of sequenced and auto calibrations.
700M HFID 6-2-15 Verifying Span Deviations → → → → → Press from the Auto Cal Deviations menu to view the Verifying Span Deviations screen. Measured Value is the averaged concentration during the Verifying Span stage of sequenced and auto calibrations.
700M HFID 6-2-15 Offset/Gain Factors → → → → When used in conjunction with the Manual Calibration Deviations, an increasing or decreasing change in Offset or Gain will provide insight into changes in analyzer performance. Press from the Calibration Factors menu to access the Offset/Gain Factors screen.
700M HFID 6-2-15 Reset Factory Settings → → → The Reset Factory Settings menu gives the operator a choice of resetting the Offsets and Gains, or both Factory Linear Coefficients and Offsets and Gains for all calibrated ranges. Resetting factory settings will not affect any other operator-changed parameters.
Page 83
700M HFID 6-2-15 Reset Offsets and Gains → → → → Pressing from the Reset Factory Settings menu will prompt the operator to confirm resetting Offsets and Gains for all ranges. Pressing (Yes) from this screen resets the Offset and Gain factors to factory default settings (0 and 1 respectively) and brings you to this confirmation screen: •...
Page 84
700M HFID 6-2-15 Reset Linear Coefficients → → → → Pressing from the Reset Factory Settings menu will prompt the operator to confirm resetting the Linear Coefficients for all ranges. Pressing (Yes) from this screen resets all the Linear Coefficients, Offset and Gain Factors and CH...
700M HFID 6-2-15 Range Setup → → Range Setup allows the operator to change Range Limits, turn Auto Range On or Off, and change Auto Range Switch Points. From the Menus screen press to access the Range Setup menu. Press to view or change Range Limits.
700M HFID 6-2-15 Range Limits → → → → The analyzer is factory-configured with four physical ranges (1 - 4). The operator can change the number of ranges and select a specific full-scale concentration in ppm. From the Range Setup menu press to access the Range Limits screen.
Page 87
700M HFID 6-2-15 4. To set two ranges, set Range 1 to the lowest value, Range 2 to the highest value, and the others to zero. 5. If new ranges are saved, the Auto Range Switch Points will be set to default percentages of range limits.
700M HFID 6-2-15 AutoRange On/Off → → → → The Auto Range Function allows the analyzer to automatically switch up and down between ranges at predetermined concentrations. From the Range Setup menu press access the Auto Range On/Off screen. The current Auto Range status appears in the upper-right corner of the screen.
700M HFID 6-2-15 AutoRange Switch Points → → → AutoRange Switch Points determine when the analyzer automatically changes a range up or down when the AutoRange function is turned on. From the Range Setup menu press to access the AutoRange Switch Points screen.
700M HFID 6-2-15 Diagnostics → → The Diagnostics menu allows the operator to access key troubleshooting screens including Diagnostic Values, Raw Values and Input/Output statuses. From the Menus screen press to access the Diagnostics menu. Press to access the Diagnostics Values screen. It allows you to check analyzer temperatures, pressures, EPC voltage percentages and flows.
700M HFID 6-2-15 Diagnostic Values → → → → Temperatures Screen The Diagnostic Values screens allow the operator to check analyzer temperatures, sample and air pressures, EPC voltage percentages and flows. These important screens are accessed by pressing from the Diagnostics menu.
Page 92
700M HFID 6-2-15 Pressures Screen The Pressures screen displays current sample and air pressures and low and high alarm limits in PSIG. EPC Voltage Percent Screen The EPC screen displays the percentage of EPC voltage being supplied to the EPC valve.
700M HFID 6-2-15 Raw Values Display → → → → The Raw Values Display screen is a diagnostic tool for viewing detector Raw Volts and Calculated Concentrations. This screen is accessed by pressing from the Diagnostics menu. Raw Voltage: This is a 0.512 VDC to 4.512 VDC that will be digitized by the microprocessor to generate the calibration curve from which the Raw Concentration and Measured Concentration are derived.
Page 94
700M HFID 6-2-15 From the Raw Values Display screen, the following functions can be useful for diagnosing and monitoring the analyzer’s performance: Press to change the analyzer’s mode to THC or CH (Methane option only). If the analyzer is in THC/CH /NMHC mode it will continue to switch between THC and CH .
700M HFID 6-2-15 I/O Status → → → The I/O Status menu gives the operator a choice of viewing the statuses of the analyzer’s digital outputs or digital inputs (open or closed). To access the I/O Status menu, press from the Diagnostics menu.
700M HFID 6-2-15 Analyzer Digital Outputs → → → → The Analyzer Digital Outputs screen allows the operator to view the status of an analyzer output (Open or Closed) and where to find the corresponding pin number. To access the Analyzer Digital Outputs status screen, press from the I/O Status menu.
Number on that connector. The abbreviations are as follows: M = Main Connector A = Auxiliary Int = internal, for CAI use only. When the analyzer is in Remote Mode and the digital input is pulled to digital ground (Main Connector, Pin 6), the status will become Closed.
700M HFID 6-2-15 Programmable Digital Outputs → → → → The Programmable Digital Outputs screen allows the operator to check the status of the analyzer’s programmable digital outputs (Open or Closed) according to pin numbers and programmed functions. From the I/O Status menu press to view Programmable Digital Output statuses.
700M HFID 6-2-15 Status Line → → → The AK Command Status Line can be displayed at the top of the Measure Screen. This field contains the current AK Protocol information. See Protocol. The current status is shown in the upper-right corner of the Status Line menu. Example above: OFF.
Digital outputs can be viewed or set up to fit the operator’s needs. Press to view or change the current TCP/IP parameters. Press to access the Data Logging Time screen. (For CAI use only) Press to view or make changes to the Auto Start Settings. Press to view or change the analyzer’s time and date.
700M HFID 6-2-15 Measure Settings → → → The Measure Settings menu provides access to the following Setup parameters: Averaging Time, THC/CH /NMHC Mode Times and CH Correction Factors. The Measure Settings menu is accessed by pressing from the Setup menu.
700M HFID 6-2-15 Averaging Time → → → → The Averaging Time screen allows the operator to set the averaging time of the measured concentration. From Measure Settings menu press to access the Averaging Time screen. The Averaging Time is a sliding average. As shown above, it can be set from 0 – 60 seconds.
700M HFID 6-2-15 THC/CH /NMHC Mode Times → → → → The THC/CH /NMHC Mode Times screen allows the operator to set the Purge Time and the Integration Time. From Measure Settings menu press to access the THC/CH /NMHC Mode Times screen.
700M HFID 6-2-15 Correction Factors → → → → → Current Settings Correction Factors allows the operator to adjust small offsets between the THC and modes while operating in the THC/CH /NMHC Mode. Press to access the CH Correction Factors menu from the Measure Settings menu.
Page 105
700M HFID 6-2-15 Press to set the analyzer in CH mode and enter the manual zero calibration screen. Introduce zero gas into the rear of the analyzer. Once the concentration stabilizes, press set the Zero Offset. Press to return to the CH Correction Factors menu.
700M HFID 6-2-15 Output Settings → → → The Output Settings menu allows the operator to change the analyzer’s Programmable Analog and Programmable Digital outputs to suit the operator’s needs. The Output Settings menu is accessed by pressing from the Setup menu.
700M HFID 6-2-15 Programmable Analogs → → → → The Programmable Analogs menu provides access to the following Setup parameters: Output Assignments, Output Scaling and Output Adjustments. The Programmable Analogs menu is accessed by pressing from the Output Settings menu.
700M HFID 6-2-15 Output Assignments → → → → → The Output Assignments screen allows the operator to view or change the signals assigned to the programmable analog outputs. From the Programmable Analogs menu press access the Output Assignments screen.
Page 109
700M HFID 6-2-15 The following output signals can be programmed by the operator using the Output Assignments screen: RealTime: In either THC or CH mode the concentration’s output will be a live reading. : In CH mode the output concentration will be a live reading. In THC/CH /NMHC mode the output will be a read and hold.
700M HFID 6-2-15 Output Scaling → → → → → The Output Scaling screen allows the operator to scale the analyzer’s Analog Outputs to a specific value for each of the four output signals. This is generally used for scaling of temperatures or pressures, but it can also be used to set an output for a specific concentration.
Page 111
700M HFID 6-2-15 EXAMPLES: 1. When the analog Output Assignment is set for Burner Temperature and the lower setting is set to 0.00 and the upper setting is set to 1000, 375°C will = 3.75 VDC. 2. When the analog...
700M HFID 6-2-15 Output Adjustments → → → → → The Output Adjustments screen allows the operator to set the output to either mA or voltage and calibrate the outputs to exactly match the results obtained on a PLC or other remote data-recording device. Using the Output Adjustments screen, the operator can force the analog outputs to 0%, 50% or 100% of full scale and back to Measure.
Page 113
700M HFID 6-2-15 Once the output type has been selected, use the Left/Right arrows to move the highlight into the % FS column of the output to be checked. Press until the % FS value reads 0.000. To adjust the zero (Offset), use the Left/Right arrows highlight the Offset column and press to open the field.
700M HFID 6-2-15 Programmable Digitals → → → → The Programmable Digitals menu provides access to the analyzer’s digital outputs for viewing and changing Output Assignments, holding or clearing alarms, and testing the outputs. The Programmable Digitals menu is accessed by pressing from the Output Settings menu.
700M HFID 6-2-15 Output Assignments → → → → → The Output Assignments screen allows the operator to assign any of the 15 programmable digital outputs to a specific alarm or status. From the Programmable Digitals menu press to access the Output Assignments screen.
Page 116
700M HFID 6-2-15 Programmable Digital Output List Alarms_________________________________________________________________ • Flame: No Flame • AIEPC: Air Inject EPC Voltage • SampP: Sample Pressure • FIEPC: Fuel Inject EPC Voltage • AirP: Air Pressure • ROvr : Over Range • FuelP: Fuel Pressure •...
Page 117
700M HFID 6-2-15 General Alarms__________________________________________________________ • Sample Pressure • Filter Temperature • Air Pressure • Sample EPC Voltage • Fuel Pressure • Air EPC Coil Voltage • Air Inject Pressure • Fuel EPC Coil Voltage • Fuel Inject Pressure • Air Inject EPC Coil Voltage •...
700M HFID 6-2-15 Output Hold/Clear → → → → → The Output Hold/Clear menu allows the operator to choose whether to hold or clear a triggered alarm when the alarm is no longer present. The current status is shown in the upper-right corner of the Output Hold/Clear menu.
700M HFID 6-2-15 Output Test → → → → → The Output Test Screen allows the operator to test the Programmable Digital Outputs to make sure they are functioning properly. The Output Test Screen is accessed by pressing from the Programmable Digitals menu. The Output Test Warning screen (shown below) first appears, asking if you wish to continue the output test.
Page 120
700M HFID 6-2-15 To test outputs one at a time, use the Up/Down arrows to highlight the desired output, then press to change the state of the output (open or closed). Press again to change the state back. To test all the outputs at once, press to change the state of all the outputs (they are all Open in the example).
The IP-address, Netmask and Gateway may be defined by the user. The Port and WinIfPort are assigned 7700 and 2000 by CAI and should not be changed unless required for a certain location. Use the Up/Down arrows to move the highlight to the setting you wish to change.
Data logging allows the analyzer to store internal variables to support CAI troubleshooting. When troubleshooting with CAI Technical Support, the operator may be asked to turn this feature on. These files can only be accessed by CAI. To access the Data Logging screen, press from the Setup menu.
700M HFID 6-2-15 Auto Start Settings → → → The Auto Start Settings screen allows the operator to set parameters that will take effect upon power up of the analyzer. The Auto Start Settings screen is accessed by pressing from the Setup menu.
Page 124
700M HFID 6-2-15 To change a setting use the Up/Down arrows to move the highlight to the setting you wish to change. Press to open the field to change the value. After making your change, press again to close the field. Press to save your changes and return to the Setup menu.
700M HFID 6-2-15 Clock Settings → → → The Clock Settings screen allows the operator to set the analyzer’s internal clock. The internal clock is used for auto calibrations and data time stamping. The Clock Settings screen is accessed by pressing from the Setup menu.
700M HFID 6-2-15 Alarms Menu → → The Alarms menu allows the operator to view Current Alarms, the Alarm Log and settable Alarm Limits. From the Menus screen press to access the Alarms menu. Press to access the Current Alarms screen and view the alarms that are currently active.
700M HFID 6-2-15 Current Alarms → → → The Current Alarms screen allows the operator to view the analyzer’s current alarms at the time this screen was accessed. To access the Current Alarms screen press from the Alarms menu. Press to refresh this screen.
700M HFID 6-2-15 Alarm Log → → → Screen Name Alarm Abbreviation Alarm Status Year / Month / Day Hour / Minute Alarm Value The Alarm Log allows the operator to view the analyzer’s last 40 alarms and their current statuses.
Page 129
700M HFID 6-2-15 Alarm Abbreviations and Descriptions Flame: No Flame FEPC: Fuel EPC Voltage SampP: Sample Pressure AIEPC: Air Inject EPC Voltage AirP: Air Pressure FIEPC: Fuel Inject EPC Voltage AInjP: Air Inject Pressure ROvr: Over Range FuelP: Fuel Pressure...
700M HFID 6-2-15 Alarm Limits → → → The Alarm Limits menu allows the operator to view or change the current upper and lower alarm tolerances. When the signals go above or lower than the assigned limit an alarm is triggered.
700M HFID 6-2-15 Temperatures → → → → From the Alarm Limits menu press to access the Temperatures screen. Use the left or right arrow to highlight the alarm limit you wish to change. Press to open the field to change the value.
700M HFID 6-2-15 Pressures → → → → From the Alarm Limits menu press to access the Pressures screen. Use the left or right arrow to highlight the alarm limit you intend to change. Press to open the field to change the value.
700M HFID 6-2-15 EPC Voltages → → → → From the Alarm Limits menu, press to access the EPC Voltage Alarm Limits screen. Use the left or right arrow to highlight the alarm limit you intend to change. Press open the field to change the value. After making a change, press again to close the field.
700M HFID 6-2-15 Concentrations → → → → Concentration alarms can be set to trigger an alarm below or above a specified concentration. From the Alarm Limits menu, press to access the Concentration Alarm Limits screen. Use the left or right arrow to highlight the concentration alarm limit you intend to change.
The scrolling alarms will be replaced with the CAI phone number when an alarm is active. If there are no alarms, only the date and time will be displayed.
6-2-15 Service Menu → → The Service Menu is for advanced operators and CAI Service. The Service menu provides access to operator-level linear coefficients and CAI factory service. From the Menus screen press to access the Service menu. Press to view or change operator-level linear coefficients.
700M HFID 6-2-15 Linear Coefficients → → → The Linear Coefficients function allows the operator to optimize linearity by inputting up to five coefficients for each range to generate up to a fourth-order curve. From the Service menu press to access the Linear Coefficients screen.
Page 138
700M HFID 6-2-15 Use the Up/Down arrows to highlight the coefficient you intend to change. Press to open the field, use the left and right arrow buttons to position the cursor, and use the Up/Down arrows or number keys to make your change. Press again to close the field.
700M HFID 6-2-15 Factory Service → → → The Factory Service menus can only be accessed by California Analytical Instruments. The best way to contact CAI with a support question is to fill out our brief Technical Support Form at http://www.gasanalyzers.com/technical_support/new.
700M HFID 6-2-15 Security → → The Security menu allows the operator to change the access to Standard or Setup Function menus and change or reset the Setup Function password. From the Menus screen press to access the Security menu.
6-2-15 Operator Levels → → → The 700M HFID has three operator access levels which allow the operator to access different analyzer functions. From the Security Menu screen press to access the Operator Levels screen. See Menu Flow Chart for security levels and functions.
700M HFID 6-2-15 Change Password → → → The Change Password screen allows the operator to change the Setup Functions password from the factory-preset 222 to a new password. From the Security menu screen press access the Operator Levels screen.
700M HFID 6-2-15 Reset Password → → → The Reset Password screen allows the operator to reset the Setup Password to the original factory password (222). From the Security menu screen press to access the Reset Password screen. To return the password to the original factory password, you must enter WIZARD in the field.
700M HFID 6-2-15 Communication and Interface Analog and Digital Interface The Main and Auxiliary connectors provide the analog outputs for concentrations and other variable signals. Digital Status outputs, Control inputs and Calibration drive signals are also provided. There are four analog outputs, whose type (mA or specific voltage range) and signal assignments are assignable from the Setup menu.
700M HFID 6-2-15 Analog and Digital Interface Hardware Capabilities of Main and Aux Connectors Analog Output The operator can choose one of the following output types: • As voltage outputs – 0 to 1V, 5V or 10V • As current outputs – 0 to 20 mA or 4 to 20 mA When set as current outputs, the maximum drive voltage provided by the analyzer is slightly more than 20 Volts, requiring that the maximum loop resistance does not exceed 1,000 Ohms.
Page 146
700M HFID 6-2-15 All opto-isolated relays are rated for 48VDC, 0.5 Amp maximum. The user is required to limit the drive current supplied to each input. All inputs are DC only and will not operate on AC current. CAUTION: Do not connect these pins directly to both sides of a voltage power supply as unrestricted current will damage the relay.
Page 147
700M HFID 6-2-15 28-Pin Main (BPM) Connector Assignments Pin # Signal Signal Type Operation Isolated Analog Ground Analog Output Isolated Analog Ground User-Defined AO-1 Analog Output 1V, 5V, 10VDC or mA User-Defined AO-2 Analog Output 1V, 5V, 10VDC or mA...
Page 148
700M HFID 6-2-15 28-Pin Auxiliary (BPA) Connector Assignments Pin # Signal Signal Type Operation Analog Ground Analog Input Analog Ground Reserved Analog Input Reserved Reserved Analog Input Reserved Reserved Analog Input Reserved Relay RTN 1 Digital Output RTN Relays 1 - 4...
700M HFID 6-2-15 Serial Interface The serial interface enables remote control of the analyzer by a master computer. It is implemented as an RS232 V24 interface and meets all requirements of the AK protocol. A 9-pin male connector at the back of the unit is used to connect a master computer with the...
Every command ends with the ETX (End of Text) character. The Error Status byte does not indicate the real number of errors. For Error Status, use the ASTF command. ® Example: Using Windows HyperTerminal for Serial RS232 Communications with CAI 700 Series Analyzers requires: 1. HyperTerminal Software 2. Windows PC/Laptop 3.
Page 151
700M HFID 6-2-15 When both HyperTerminal and the analyzer are running while connected by the null modem cable, the analyzer will present a menu if sent a non-AK command. This menu can be ignored and AK commands will be acted on by the analyzer. The picture below shows a sequence with the AKON 0 command being sent to the analyzer.
700M HFID 6-2-15 Instruction Command Byte Character Explanation Byte ASCII Code 02 Byte Don’t Care Any ASCII code Byte Function Code 1 AK instruction, e.g. ASTF Byte Function Code 2 AK instruction, e.g. ASTF Byte Function Code 3 AK instruction, e.g. ASTF...
Page 153
700M HFID 6-2-15 Acknowledgement Command Byte Character Explanation Byte ASCII Code 02 Byte Don’t Care Any ASCII code Byte Function Code 1 Echo of the AK Instruction Command Byte Function Code 2 Echo of the AK Instruction Command Byte Function Code 3...
700M HFID 6-2-15 Error Handling It is possible that an unknown instruction is sent, the analyzer is busy with a function that is not the desired one, or an error occurred in the command parameters. The table below provides a summary of all errors that can appear upon any master instruction.
700M HFID 6-2-15 General AK Requirements 1. If the command message contains no error, the Acknowledge message contains the echo of the Function code and the Error Status number (0 to 9). 2. If the transfer was faulty or the function code is unknown, the answer contains four question marks (for example, ???? 0).
Page 156
700M HFID 6-2-15 Abbreviations Abbreviation Description Measuring Range Number Mn .. M4 Measuring Range 1 .. 4 w.w .. ZZ. Numerical Value Number Numeric Integer Value a0 .. a4 Polynomial Coefficients Status Yyymmdd Date of Format Year, Month and Day with 2 characters each and no spaces...
700M HFID 6-2-15 Scan Commands AKON: Measured Concentration Value Command Response Description _AKON_K0 _AKON_s_z.z_y.y_x.x_t z.z : Current Measure Value y.y : CH concentration x.x : NMHC concentration w.w : THC concentration Note: y.y, x.x, and w.w are normally 0.0 when “THC/CH...
Page 158
700M HFID 6-2-15 AMBE: Measuring Range Limit Command Response Description _AMBE_K0 _AEMB_s_M1_w.w All existing measuring range limits. _M2_x.x _M3_y.y _M4_z.z AKAK: Calibration Gas Concentrations Command Response Description _AKAK_K0 _AKAK_s_M1_w.w All existing calibration gas values are responded. _M2_x.x _M3_y.y _M4_z.z _AKAK_s_Mn_z.z Calibration gas value of Range n.
Page 159
700M HFID 6-2-15 ASTZ: Normal Device Status Command Response Description _ASTZ_K0 _ASTZ_s_State 1_State 2_. . State 4 Respond device status for all channels. Possible States Possible States Response Description SREM Remote State 1 SMAN Manual State 2 STBY Standby SPAU...
Page 160
700M HFID 6-2-15 ASTF: Error Status Command Response Description _ASTF_K0 _ASTF_s_f1_f2_f3_...f26 Current error numbers of all are responded. Error Numbers and Descriptions No Flame Check Sample Pressure Check Air Pressure Check Fuel Pressure Check Air Inject Pressure Check Fuel Inject Pressure...
Page 161
700M HFID 6-2-15 AKEN: Device Identification Command Response Description _AKEN_K0 _AKEN_s_devicename Device identification is responded. _AKEN_K1 _AKEN_s_model Device Model _AKEN_s_serial no Device Serial Number _AKEN_K2 _AKEN_K3 _AKEN_s_Air pressure Air Pressure _AKEN_s_Sample pressure Sample Pressure _AKEN_K4 _AKEN_K5 _AKEN_s_Fuel pressure Fuel pressure...
Page 162
700M HFID 6-2-15 ADUF: Flows Command Response Description _ADUF_K0 _ADUF_s_z.z_y.y 1. Sample Flow 2. Air Flow 3. Fuel Flow _ADUF_K0_x _ADUF_S_xflow Flow of x AGRD: Polynomial Coefficients Command Response Description _AGRD_s_a0_a1_a2_a3_a4 Polynomial coefficients of Channel m Range n _AGRD_K0_Mn AKAL: Percent Deviations of Last Accepted Calibration...
Page 163
700M HFID 6-2-15 AAEG: Verifying Span-Point Deviation During Auto Calibration Command Response Description _AAEG_K0 _AAEG_s_M1_z.z_da_dr Verifying deviations of Ranges M1 to M4 _M2_z.z_da_dr from span point stored after auto calibration. _M3_z.z_da_dr Values: measured value (z.z), absolute dev _M4_z.z_da_dr (da), relative dev (dr).
Page 164
700M HFID 6-2-15 APAR: Auto Calibration Tolerance Values Command Response Description _APAR_K0_SATK _APAR_s_z.z_y.y_x.x_w.w Auto calibration tolerance value (%): z.z: Range 1 y.y: Range 2 x.x: Range 3 w.w: Range 4 ASYZ: Respond System Time Command Response Description _ASYZ_s_yymmdd_hhmmss System Time:...
Page 165
700M HFID 6-2-15 ADAL: Diagnostic Alarm Limits Command Response Description _ADAL_K0 _ADAL_s_a1.min_a1.max_..All alarm limits. _a16.min_a16max _ADAL_s_x.min_x.max Alarm limits of x _ ADAL_K0_x Alarm Limit Numbers and Descriptions Sample Pressure Air Pressure Fuel Pressure Air Inject Pressure Fuel Inject Pressure...
700M HFID 6-2-15 Control Commands SARA: Auto Range Off Command Response Description _SARA_s Set auto range Off. _SARA_K0 SARE: Auto Range On Command Response Description _SARE_K0 _SARE_s Set auto range On. SATK: Start Automatic Calibration Command Response Description _SATK_K0 _SATK_s Start automatic calibration of all available ranges.
Page 169
700M HFID 6-2-15 SEMB: Set Measuring Range Command Response Description _SEMB_K0_Mn _SEMB_s Set measuring range to range ‘ n ‘. Auto range will be disabled. SENT: Set Calibration Gas Flow (Pumps or Valves) Command Response Description _SENT_K0_X _SENT_s x: 10 = Pump...
Page 170
700M HFID 6-2-15 SNMH: Sets THC/CH4/NMHC Mode Command Response Description _SNMH_K0 _SNMH_s Sets the measurement mode to THC/CH4/NMHC. Automatic switching between THC and CH4 modes SPAU: Pause Command Response Description _SPAU_s Pause mode (Shuts off Flame) _SPAU_K0 SRES: Reset Command...
Page 171
700M HFID 6-2-15 SUDP: Start/Stop UDP Data Streaming Command Response Description _SUDP_K0_ON _SUDP_s Start data streaming via the UDP channel. Configure the channel before starting with EUDP command. _ SUDP_K0_OFF _SUDP_s Stop streaming via the UDP channel. SVZS: Reset Offset to 0 and Gain to 1...
700M HFID 6-2-15 Configuration Commands EKAK: Set the Four Span Gas Concentration Values Command Response Description _EKAK_K0_M1_ _EKAK_s Set the span gas values. w.w_M2_x.x_M3_ y.y_M4_z.z EMBE: Set the Four Measuring Range Full Scale Limits Command Response Description _EMBE_K0_M1_ _EMBE_s Set the range full scale limits.
Page 173
700M HFID 6-2-15 EFDA: Set Auto Calibration and Purge Times Command Response Description _EFDA_K0_SATK_ _EFDA_s Set auto calibration times: z_y_x z: Purge time y: Verify time x: Purge after (z, y, x, w in seconds) _ EFDA_K0_ _EFDA_s Set analyzer purge time to z seconds.
Page 174
700M HFID 6-2-15 EDAL: Set Diagnostic Alarm Limits Command Response Description _EDAL_K0_ai.min_ _EDAL_s Set all alarm limits. a1.max_..._a16max _EDAL_s Set alarm limit of x. _EDAL_K0_x_ x.min_xmax Alarm Numbers and Descriptions Sample Pressure Air Pressure Fuel Pressure Air Inject Pressure Fuel Inject Pressure...
Page 175
700M HFID 6-2-15 ETCP: Set TCP/IP Parameters Command Response Description _ETCP_K0_ _ETCP_s Zzz = TCP/IP address zzz.zzz.zzz.zzz_ yyy = TCP/IP subnet mask xxxx = TCP/IP port yyy.yyy.yyy.yyy_xxxx All changes take effect after next Power On cycle. EUDP: Set TCP/IP Data Streaming Parameters...
Page 176
700M HFID 6-2-15 EATK: Set Auto Calibration Parameters Command Response Description _EATK_K0_z_y_x _EATK_s_z_y z: 1) CH Mode 2) THC Mode y: 1) ALL gases 2) Zero gas only EFDA: Set Auto Calibration Time Command Response Description _EFDA_K0_SATK_z_y_x _EFDA_s_z_y_x z: Purge time...
700M HFID 6-2-15 Modbus Protocol Modbus on TCP/IP Application Data Unit MBAP Description This section describes the encapsulation of a Modbus request or a response when it is carried on a Modbus TCP/IP network. A dedicated header, called the MBAP (Modbus Application Protocol) header, is used on TCP/IP to identify the Modbus Application Data Unit.
All Modbus/TCP ADU are sent via TCP to Registered Port 502. The different fields are encoded in Big-endian. The 700M HFID analyzer uses only the Length bytes from the MBAP section.
700M HFID 6-2-15 Modbus Command Function Codes Code 01 This function code is used to read from 1 to 2000 contiguous status bits in a remote device. The requesting remote device specifies the starting address, including the address of the first bit specified and the number of bits.
Page 180
700M HFID 6-2-15 Error Function Code 1 Byte Function code + 0x80 Exception Code 1 Byte 01 or 02 or 03 or 04 Here is an example of a request to read discrete outputs 20–38: Request Response Field Name (Hex)
Page 182
700M HFID 6-2-15 Code 03 This command has been modified to read floating-point numbers in 32-bit IEEE format. This function code is used to read the contents of a contiguous block of floating-point registers in a remote device. The Request PDU specifies the starting register address and the number of registers.
Page 183
700M HFID 6-2-15 Here is an example of a request to read Register 0: Request Response Field Name (Hex) Field Name (Hex) Function Function Starting Address Hi Byte Count Starting Address Lo Register Value Hi (1) Number of Registers Hi...
Page 184
700M HFID 6-2-15 Here is an example of a request to read three registers starting at 40201: Request Response Field Name (Hex) Field Name (Hex) Function Function Starting Address Hi Byte Count Starting Address Lo Register Value Hi (40202) Number of Registers Hi...
Page 185
700M HFID 6-2-15 Command 03, Read four Floating Point values starting at Address 1. Result = 1234.56789, 0.0 -1234.568, 10000.
Page 186
700M HFID 6-2-15 Code 04 This function code is used to read from 1 to 125 contiguous input registers in a remote device. The Request PDU specifies the starting register address and the number of registers. In the PDU, registers are addressed starting at zero. Therefore input registers numbered 1-16 are addressed as 0-15.
Page 187
700M HFID 6-2-15 Here is an example of a request to read Input Register 8: Request Response Field Name (Hex) Field Name (Hex) Function Function Starting Address Hi Byte Count Starting Address Lo Input Register 9 Hi Quantity of Input Registers Hi...
Page 188
700M HFID 6-2-15 Code 05 This function code is used to write a single output to either ON or OFF in a remote device. The requested ON/OFF state is specified by a constant in the request data field. A value of FF 00 hex requests the output to be ON.
Page 189
700M HFID 6-2-15 Here is an example of a request to write bit 173 ON: Request Response Field Name (Hex) Field Name (Hex) Function Function Output Address Hi Output Address Hi Output Address Lo Output Address Lo Output Value Hi...
Page 190
700M HFID 6-2-15 Code 06 This function code is used to write a single holding register in a remote device. The Request PDU specifies the address of the register to be written. Registers are addressed starting at zero. Therefore the register numbered 1 is addressed as 0.
Page 191
700M HFID 6-2-15 Here is an example of a request to write Register 1 to 00 03 hex: Request Response Field Name (Hex) Field Name (Hex) Function Function Register Address Hi Register Address Hi Register Address Lo Register Address Lo...
Page 192
700M HFID 6-2-15 Code 16 This function code is used to write a single floating point register to a remote device. The requested written values are specified in the request data field. Data is packed as four bytes per register. The normal response returns the function code, starting address and quantity of registers written.
Page 193
700M HFID 6-2-15 Here is an example of a request to write two registers starting at 2 to 00 0A and 01 02 hex: Request Response Field Name (Hex) Field Name (Hex) Function Function Starting Address Hi Starting Address Hi...
Page 194
700M HFID 6-2-15 Command 16, Write one floating point register at Address 40200 showing error response. Not a valid address.
Page 195
700M HFID 6-2-15 Code 26 This is a non-standard code used to read an ASCII string. Request Function Code 1 Byte 0x1A Starting Address 2 Bytes 0x0000 to 0xFFFF Quantity of Registers 2 Bytes Response Function Code 1 Byte 0x1A...
700M HFID 6-2-15 Modbus Map 01H Single-Read Coil Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.) Coil Numbers and Descriptions Coil Number Read Data No Flame Sample Pressure Air Temperature Fuel Temperature Air Inject Pressure (if fitted)
Page 197
700M HFID 6-2-15 Coil Number Read Data High Concentration 0 - Standby, 1 - Measure General Alarm Eng units For additional alarms and status 0 - Manual, 1 - Remote 0 - Standby, 1 - Measure 1 - Zero, 1 - Span,...
700M HFID 6-2-15 05H Write Single Coil Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.) Coil Numbers and Descriptions Coil Number Write Data 0 - Manual, 1 - Remote 0 - Standby, 1 - Measure...
700M HFID 6-2-15 03H Read Floating Point Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.) Register Numbers and Descriptions Register Number Contents IEEE 40001 UNDILUTED Real Time Concentration = Diluted Conc* Dil. Ratio / 10000...
Page 200
700M HFID 6-2-15 Register Number Contents IEEE 40063 Range 1 Gain 40065 Range 2 Offset 40067 Range 2 Gain 40069 Range 3 Offset 40071 Range 3 Gain 40073 Range 4 Offset 40075 Range 4 Gain 40109 Range 1 Full Scale...
Page 201
700M HFID 6-2-15 Register Number Contents IEEE 40239 Air inject Pressure Minimum (if fitted) 40241 Air inject Pressure Maximum (if fitted) 40243 Fuel Inject Pressure Minimum (if fitted) 40245 Fuel Inject Pressure Maximum (if fitted) 40247 Filter Temperature Minimum 40249...
700M HFID 6-2-15 16H Write Floating Point Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.) Register Numbers and Descriptions Register Number Contents IEEE 40201 Range 1 Span Gas Concentration 40203 Range 2 Span Gas Concentration...
Page 203
700M HFID 6-2-15 Register Number Contents IEEE 40267 Sample EPC Voltage Minimum 40269 Sample EPC Voltage Maximum 40271 Air EPC Voltage Minimum 40273 Air EPC Voltage Maximum 40275 Fuel EPC Voltage Minimum 40277 Fuel EPC Voltage Maximum 40279 Air Inject EPC Voltage Minimum (if fitted)
Buyer by CAI and if any such product should prove to be defective within such one year period, CAI agrees, at its option, either (i) to correct by repair or, at CAI’s election, by replacement with...
Page 205
700M HFID 6-2-15 f) CAI may from time to time provide a special printed warranty with respect to a certain product, and where applicable, such warranty shall be deemed incorporated herein by reference; g) CAI shall be released from all obligations under all warranties, either expressed or implied, if any product covered hereby is repaired or modified by persons other than its own authorized service personnel unless such repair by others is made with the written consent of CAI.
Need help?
Do you have a question about the 700M and is the answer not in the manual?
Questions and answers