HART® protocol
62
No and function
34
Write damping value for the PV
35
Write range values for the PV
38
Reset "config changed" flag
40
Enter/exit fixed current mode loop
test (fix the analogue current at
specified value
128
Read emissivity
129
Write emissivity
130
Read reflectivity
131
Write reflectivity
132
Read transmissivity
133
Write transmissivity
134
Read error current
135
Write error current
136
Read max/min hold time
137
Write max/min hold time
138
Read alarm values
139
Write alarm values
144
Special command (read only)
see description
145
Special command
see description
146
Read factory settings and write
them to EEPROM
Data in the instruction
float
average value 0...999.9 s
byte
range unit code
float
upper range value (MBE)
float
lower range value (MBA)
float
current
(0 = exit fixed current mode)
float
Emissivity 0.100...1.000
float
Reflectivity 0.100...1.000
float
Transmissivity 0.100...1.000
float
error current
float
Max hold time 0.0 ... 999.9 s
float
Min hold time 0.0 ... 999.9 s
float
Alarm value [°C]
float
Value 1
float
Value 2
float
Value 1
float
Value 2
int
Password
Password [32] >only RAM
Password [34] >EEPROM
Operating Instructions Radiation sensors R3XX
Data in the reply
float
average value 0...999.9 s
byte
range unit code
float
upper range value (MBE)
float
lower range value (MBA)
float
current
float
Emissivity
float
Emissivity
float
Reflectivity
float
Reflectivity
float
Transmissivity
float
Transmissivity
float
error current [mA]
float
error current [mA]
float
Max hold time [s]
float
Min hold time [s]
float
Max hold time [s]
float
Min hold time [s]
float
Alarm value [°C]
float
Alarm value [°C]
float
Value 1
float
Value 2
float
Value 1
float
Value 2
int
Success message (0 = error ; 1 =
ok)

Hide quick links: