15. MEASURED VALUE STORAGE

All measurements performed are saved automatically. All values can be read again and deleted in this menu.
To call the saved value rate settings, push the Menu button (3); then use the navigation buttons "r" and "s" (2) to
select the measured value storage (folder icon at the left of the display).
Now push the "OK" button (4) to get access to the different saved values. After pushing the "OK" button (4), the
folder icon will start flashing; the navigation buttons "r" and "s" (2) can be used to display the different memory
contents.
If no saved values are present, there will be four dashes shown in the display.
Display of the radiation load per hour
A value is only indicated here if any measurement has been active for at least one minute and has been completed.
This also applies if the measurement is cancelled.
The "Gamma Check Pro" shows an accumulated value of the measured radiation dose for the radiation load
per hour. The incoming radiation is measured for a period of at least 60 seconds and extrapolated to the
value per hour.
You will receive a value here as soon as the measurement has continued for at least one hour. When meas-
uring for several hours, the radiation dose will be updated automatically if it has increased.
Day dosage value
The average value of the last 24 hours is displayed here; this therefore corresponds to a "day dosage value", i.e. the
total radiation that one has been exposed to in the last 24 hours.
A value is only indicated here if any measurement has been active for at least 24 hours. Even if a measurement has
been performed for several days, only the average of the last 24 hours is displayed here.
24h
50
µSv
h
MAX
µSv
h