6.5 Battery Disassembly

This section describes the steps to remove the battery from the N5500 for replacement
without disassembling the main case of the monitor
R. Battery Disassembly
N5500 Service Manual
Disassembly Guide
a. Remove 2 screws on the lower part of the
rear case, holding the Battery cover to the
rear case.
b. Remove the Battery cover.
c. Disconnect the Battery wire.
d. Remove the Battery.