Wallach Colposcopes EMC Compliance Information

MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in
the ACCOMPANYING DOCUMENTS.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Guidance and Manufacturer's Declaration – Electromagnetic
Emissions
The Wallach Colposcope is intended for use in the electromagnet environment specified below.
The customer or the end user of the Wallach Colposcope should assure that it is used in such an
environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
IMSC036-DFU • Rev. B • 10/14
Compliance
Electromagnetic Enviroment- Guidance
Wallach Colposcopes use RF energy only for it internal function.
Therefore, its RF emissions are very low and are not likely to
Group 1
cause any interference in nearby electronic equipment.
Wallach Colposcopes are suitable for use in all establishments,
including domestic establishments and those directly connected
Class A
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class A
Complies
16
Made in the USA