Electrical

The Wallach Colposcope with Trulight requires a 100-240V ~50-60 Hz power source to
operate. It is supplied through a 2.5 meter medical grade detachable cord. For your safety,
assure that your electrical outlets meet code regulations.

Replacing the Fuses

1. Shut OFF (labeled "O") the master switch and unplug the power cord.
2. The electronics box contains two fuses beneath the power socket.
3. To remove, pull out the drawer beneath the power socket.
4. Remove the blown fuse from the holder and replace with a 2 amp, 5mm x 20mm Fast Blow
Fuse (Part Number 109645).
5. Push the fuse holder back into the electronics box.

Maintenance

PRECAUTION: DO NOT IMMERSE EQUIPMENT COMPONENTS IN LIQUID; UNPLUG
BEFORE CLEANING; DRY BEFORE USE.
Clean the lenses and protective light covers with silicone-treated lens tissue paper ONLY.
The colposcope finish is a high-gloss enamel and may be wiped down with disinfectant. Keep the
instrument clean and dust-free. When the colposcope is not in use, replace protective covers.

Warranty Information

The Wallach Star series of colposcopes is supported by a seven-year warranty from date of
purchase covering any failure of the device due to defective workmanship or components, when
used in compliance with the product's intended use.