Periodic Maintenance; Cleaning Procedure; Troubleshooting - Hanna Instruments Gro Line HI9814 Instruction Manual

Ph, ec, tds and temperature meter with quick calibration
Table of Contents

Advertisement

STORAGE
• To minimize clogging and ensure a quick response time,
the glass bulb and the junction should be kept moist and
not allowed to dry.
• Replace protective cap with a few drops of
Storage Solution.
• Follow PREPARATION above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED WATER.

PERIODIC MAINTENANCE

• INSPECT the probe especially the pH bulb for any
scratches or cracks. If any present, replace the probe.

CLEANING PROCEDURE

• Soak in Hanna
HI700661P
general purpose cleaning solution for
agriculture for 20 minutes. Rinse well and calibrate
before using.
IMPORTANT: After performing any of the cleaning procedures
rinse the electrode thoroughly with distilled water. Soak
the electrode in
HI70300
Rinse well. All storage solution must be removed from the
EC probes to function properly. The storage solution is highly
conductive. Calibrate probe before taking measurements.

TROUBLESHOOTING

Meter: Follow operating and calibration procedures from
this instruction manual.
pH Electrode: Evaluate your pH electrodes performance by
performing the SENSOR CHECK procedure found on page 13.
EC Cell: Rinse probe tip with copious amounts of tap water.
Then rinse with deionized or distilled water. Shake off water
droplets. Place tip in a fresh sample of conductivity (or TDS)
standard. Observe reading on display. If the reading is close
to standards value, a calibration should correct the value.
If the value is drifting, inspect to see if matter is caught
around the cell electrodes. Remove matter carefully, rinse
well and repeat test. If the EC reading continues to drift,
the probe should be replaced.
HI7061
general cleaning solution or
Storage Solution for 30 minutes.
HI70300
17