IX. Advanced Function Setting
PressTARE until the screen shows sensor code, then press
MODE will enter the advanced function setting (Press CAL
to select, press MODE to confirm, press ON/OFF to exit.).
1) C Function (Zero Tracking Selection)
C---1: No zero tracking
C---2: Zero tracking +/-1d (default)
C---3: Zero tracking +/-2d
Note: The value is larger, the system is more stable; the
value is smaller, the system is more sensitive.
2) P Function (Data Transmission Mode Selection)
P---0: stop transmission.
P---1: Single transmission, data transmission according to
press PRINT once.
P---2: Continuous transmission. Press PRINT to transmit
continuously, and then click PRINT to stop continuous transmission.
P---3: Single transmission when stable. When the weight is
stable, it'll print once. (Default)
3) CLA Function (transmission data)
CLA--0: transfer current scale weight. (default)
CLA--1: transfer series number, net weight, tare weight, gross
weight.
4) FA Function (weighing speed)
FA---1: Weighing speed is natural. (default)
FA---2: Weighing speed is faster.
Note: After changing the weighing speed, please calibrate the
balance.
5) b Function (Baud Rate Selection)
b---1: 1200 bps
b---2: 2400 bps