Directions for use
Transducer accuracy test
Required tools, equipment, and accessories
To complete the transducer accuracy test, the following tools and accessories are
required:
•
scissors or other cutting device
•
one 4-way hose barb tee for 1/8 inch ID tubing
•
a minimum of 32 inches of 1/4 inch (.25) OD and 1/8 inch (.125) ID silicone rubber
tubing
○
○
○
For the transducer accuracy test, the following equipment is required: (1) the test
volume repair fixture with test manifold, bulb, and valve; (2) a pressure meter simulator;
(3) the device with the Flexiport
the test equipment, contact Welch Allyn Technical Support at: http://
www.welchallyn.com/support.
No.
1
2
3
1. Use scissors or other cutting device to cut off the Flexiport
end of the device tubing.
approximately 14 inches from the hand bulb to the 4-way tee
approximately 12 inches from the test volume repair fixture to the 4-way tee
approximately 6 inches from pressure meter simulator to the 4-way tee
®
connector removed. For further information or to order
Item
Test volume repair fixture
(Welch Allyn BP Test Volume Repair Fixture 407672 or equivalent
with test manifold, bulb and valve, 4-way hose barb tee, and 1/4
inch [.25] OD silicone rubber tubing)
Note
For the transducer accuracy test, connect the
test equipment shown below. Connect the
manifold to the test volume repair fixture. The
calibrated volume is 500 ml.
Pressure meter simulator (with an accuracy of ± 3 mmHg)
Device
(ProBP™ 2000 Digital BP Device)
Specifications 19
®
hose fitting from the