DeVilbiss Drive AirForce Mini User Manual page 36

Micro-compressor portable nebulizer
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 19
EN
Guidance and manufacturer's declaration - electromagnetic emissions
The
A
irFo ce Mini
r
mer or the user of
Immunity test
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power fre-
quency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mians voltage prior to application of the test level.
34
is intended for use in the electromagnetic environment specified below. The custo-
t
his
device
should assure that it is used in such an environment
IEC 60601 Test level
Power supply
lines :±2 kV
input/output lines:±1 kV
line(s) to line(s): ±1 kV
line(s) to earth: ±2 kV
100 kHz repetition
frequency
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º,
180 º, 225 º, 270 º and
315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
3 A/m
50Hz/60Hz
Compliance level
Power supply
lines :±2 kV
input/output lines:±1 kV
line(s) to line(s): ±1 kV
to line(s) to earth:±2 kV
100 kHz repetition
frequency
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º,
180 º, 225 º, 270 º and
315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
3 A/m
50Hz/60Hz
Electromagnetic
environment - gui-
dance
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency ma-
gnetic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mq6010

Table of Contents