laerdal LSR Adult User Manual

Silicone resuscitator

Advertisement

Quick Links

EN
Laerdal Silicone
Resuscitators
User Guide
www.laerdal.com

Advertisement

Table of Contents
loading

Summary of Contents for laerdal LSR Adult

  • Page 1 Laerdal Silicone Resuscitators User Guide www.laerdal.com...
  • Page 2: Table Of Contents

    Contents Indication for Use Intended Use Important Information Items Included Adult Model Overview Adult Model Assembly and Disassembly Paediatric Model Overview Paediatric Model Assembly and Disassembly Preterm Model Overview Preterm Model Assembly and Disassembly Intake Valve Assembly and Disassembly Intake Valve Cap Update Function Test Cleaning and Disinfection Regulatory Information...
  • Page 3: Indication For Use

    The Preterm model is intended for patients below 2.5 kg (5.5 lb). order of a physician. • The LSR should only be used by persons who have received This User Guide applies to all three models of the Laerdal Silicone adequate training in the use of resuscitators. Resuscitator unless otherwise specified.
  • Page 4: Items Included

    Items Included Items Included Caution Paediatric Model Do not use parts other than genuine Laerdal parts. Use of non-Laerdal (Cat. No. 86xxxx) parts may affect safety and/or performance. Items may vary in appearance and are subject to change. Adult Model...
  • Page 5: Adult Model Overview

    Adult Model - Overview Adult Model - Overview Reservoir Flap Valves (x2) Lip Valve Patient Valve Disk Membrane Intake Valve Reusable Oxygen Oxygen Port Adult Ventilation Bag Reservoir Bag Silicone Mask Intake Disk Membrane with Multi-Function Mask Cover...
  • Page 6 Adult Model - Overview Adult Model - Overview Assembly and Disassembly C L I C K Cautions Improper assembly of the flap valves, intake membrane, disk membrane and lip valve may affect performance. Ensure use of one lip valve. Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure (PEEP) or prevent proper patient exhalation.
  • Page 7: Paediatric Model Overview

    Paediatric Model - Overview Paediatric Model Overview Reservoir Flap Valves (x2) Pressure Relief Valve Lip Valve Patient Valve Disk Membrane Intake Valve Paediatric Ventilation Bag Oxygen Port Reusable Oxygen Silicone Mask Intake Disk Membrane Reservoir Bag (Child 3-4) with Multi-Function Mask Cover...
  • Page 8 Paediatric Model - Overview Paediatric Model Overview Assembly and Disassembly C L I C K Caution Improper assembly of the flap valves, intake membrane, disk membrane and lip valve may affect performance. Ensure use of one lip valve. Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure (PEEP) or prevent proper patient exhalation.
  • Page 9: Preterm Model Overview

    Preterm Model - Overview Preterm Model - Overview Reservoir Flap Valves (x2) Pressure Relief Valve Lip Valve Patient Valve Disk Membrane Intake Valve Preterm Ventilation Oxygen Port Silicone Mask Reusable Oxygen Intake Disk Membrane Reservoir Bag...
  • Page 10 Preterm Model - Overview Preterm Model - Overview Assembly and Disassembly C L I C K Caution Improper assembly of the flap valves, intake membrane, disk membrane and lip valve may affect performance. Ensure use of one lip valve. Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure (PEEP) or prevent proper patient exhalation.
  • Page 11: Intake Valve Assembly And Disassembly

    Intake Valve Assembly/Disassembly Intake Valve Cap Update Click on - click o New version of cap and valve are indentified by textured circle. Post 2015 cap C L I C K Pre 2015 cap Post 2015 valve Post 2015 cap Pre 2015 cap Pre 2015 valve...
  • Page 12: Function Test

    Squeeze Seal neck Caution If a Laerdal Silicone Resuscitator fails any of these function tests it is to be correctly cleaned, re-assembled and tested or removed from service and not used. Release bag Inspect Bag - must...
  • Page 13 Function Test Function Test Testing the Patient Valve Testing the Disk Membrane of Patient Valve Caution Ensure that a (single) Lip Valve has been installed in the Patient Valve. Fill reservoir bag with air Attach reservoir bag to Patient Valve Squeeze Inspect Disk Membrane...
  • Page 14 Function Test Function Test Testing the Outlet Reservoir Flap Valve Testing the Inlet Reservoir Flap Valve with Ambient Air Attach reservoir Fill reservoir Attach reservoir bag to Fill bag to bag with air Patient Valve reservoir Patient Valve bag with Attach reservoir bag to Intake Valve Attach reservoir bag to Intake Valve Seal oxygen port and...
  • Page 15: Cleaning And Disinfection

    Ensure Pressure Relief Valve is functioning prior to use. malfunction. Laerdal is not responsible for any consequences of re-use. Cautions • Laerdal strongly discourages the use of rinsing and drying agents. Such agents may not be compatible with the materials used in the Laerdal Silicone Resuscitator.
  • Page 16 Cleaning and Disinfection Cleaning and Disinfection Washing and Rinsing To reduce the risk of cross-contamination, follow these instructions after each use. The LSR must be cleaned before high-level disinfection or sterilisation. The LSR can be manually cleaned, or cleaned in an automatic Inspection washer/disinfector.
  • Page 17: Regulatory Information

    Council Directive 93/42/EEC as amended 2460 by Council Directive 2007/47/EC. Sterilisation/High-level Disinfection Method Process Parameters Post-Treatment Laerdal Silicone Resuscitator meets the following Standards: Temperature / Exposure Concentration time • EN 1789 • ISO 10651-4 Sterilisation (applies to all parts except the Reservoir Bags) •...
  • Page 18: Specifications

    Specifications Specifications Preterm Model Not made with natural rubber latex. Ventilation Bag volume: 240 ml Adult Model Reservoir Bag volume: 600 ml Ventilation Bag volume: 1600 ml Delivered O concentrations under various test conditions Reservoir Bag volume: 2600 ml flow Tidal volume (ml) x bag cycling rate per minute.
  • Page 19 LSR Head strap Useful Life 870120 LSR Hanging loop Laerdal Silicone Resuscitators should undergo the Function Test 54010733 High Efficiency Filter for LSR/TheBAG(IE)(25pk) at least once every year. If a correctly assembled LSR fails the test, the failing parts must be replaced and disposed of.
  • Page 20 Cap for LSR Intake Valve spare part pkg. 3 872220 Adult & Child Sil. Mask w/Multi Func. Mask Covers 860420 LSR Compact Case Adult For latest version of Spare Parts and Accessories, visit www.laerdal.com. 870600 LSR Display case cpl. Adult 860300 Display Case, Paediatric Warranty Refer to the Laerdal Global Warranty for terms and conditions.
  • Page 21 ® Laerdal is a registered trademark of Laerdal Medical AS. © 2017 Laerdal Medical AS. All rights reserved. Manufactured in China for Laerdal Medical AS P.O. Box 377, Tanke Svilandsgate 30, 4002 Stavanger, Norway T: (+47) 51 51 17 00...

This manual is also suitable for:

Lsr paediatricLsr preterm

Table of Contents