Specifications - KAM Simple Precision PKF User Manual

Karl fischer moisture analyzer
Table of Contents

Advertisement

2. SPECIFICATIONS

Method:
Detection:
Control:
Display:
Sample size:
Range:
Sensitivity:
Accuracy:
Generator electrode config.:
Titration speed:
Power requirements:
Ambient Temperature
Communication:
Printer:
Dimensions:
Weight:
KFMANUAL 0317
Coulometric Karl Fischer titration
Polarization detection
Automatic electrolysis current control
320 x 240 pixel LCD
0.1, 0.25, 0.5, 1.0 ml
or less than 2 grams (or ml)
10 μg - 100,000 μg H
1 μg H
O
2
±5 μg for 10 μg - 1000 μg,
0.5% (C.V.) for over 1000 μg
(meets or exceeds API MPMS 10.9, ASTM D4928, ISO 12937 ,
and EI 386)
With diaphragm
1000 μg H
O/min. (max. at high H
2
Operates on either AC or DC.
AC - 110/120, 220/240 V, 50/60 Hz
12 V rechargeable lithium battery
5ºC – 40º C
USB port, BlueTooth
Optional
Portable 15.5" x 8"x 10" (394 mm x 203 mm x 254 mm)
Lab 10.5" x 8" x 9"(267 mm x 203 mm x 229 mm)
Portable approx. 20 lbs. (9 kg)
Lab approx. 10 lbs. (4.5 kg)
O
2
O concentrations)
2
, Windows
based software
®
®
4
KAM CONTROLS, INC.

Advertisement