Power; Table 2 Table Of Various Bath Fluids - Fluke 7380 User Manual

Ultra low temp bath
Table of Contents

Advertisement

Table 2 Table of Various Bath Fluids

Fluid
Lower
(# = Hart Part No.)
Temperature
Limit*
Halocarbon 0.8
–90°C (v)**
#5019
Methanol
–96°C (fr)
Water
0°C (fr)
Ethylene
–35°C (fr)
Glycol—50% #5020
Mineral Oil
40°C (v)
Dow Corning 200.5
–40°C (v)**
Silicone Oil
Dow Corning 200.10
–35°C (v)**
#5012
Dow Corning 200.20
7°C (v)
#5013
Dow Corning 200.50
25°C (v)
Silicone Oil
Dow Corning 550
70°C (v)
#5016
Dow Corning 710
80°C (v)
#5017
Dow Corning 210-H
66°C (v)
Silicone Oil
Heat Transfer Salt
145°C (fr)
#5001
*Limiting Factors — b - boiling point
**Very low water solubility, ice will form as a slush from condensation below freezing.
8.3
Upper
Temperature
Flash
Viscosity
Limit*
Point
(centistokes)
70°C (e)
NONE
5.7 @ –50°C
0.8 @ 40°C
0.5 @ 70°C
60°C (b)
54°C
1.3 @ –35°C
0.66 @ 0°C
0.45 @ 20°C
95°C (b)
NONE
1 @ 25°C
0.4 @ 75°C
110°C (b)
NONE
7 @ 0°C
2 @ 50°C
0.7 @ 100°C
190°C (fl)
190°C
15 @ 75°C
5 @ 125°C
133°C (fl, cc)
133°C
5 @ 25°C
165°C (fl, cc)
165°C
10 @ 25°C
3 @ 135°C
230°C (fl, cc)
230°C
20 @ 25°C
280°C (fl, cc)
280°C
50 @ 25°C
232°C (fl, cc)
232°C
50 @ 70°C
300°C (fl, oc)
10 @ 104°C
302°C (fl, oc)
302°C
50 @ 80°C
7 @ 204°C
315°C (fl, oc)
315°C
50 @ 66°C
14 @ 204°C
530°C
NONE
34 @ 150°C
6.5 @ 300°C
2.4 @ 500°C
e - high evaporation
fl - flash point

Power

Power to the bath is provided by an AC mains supply. See Section 3.1, Specifi-
cations, for power details. Refer to and read the CAUTION at the front of this
manual concerning brownout and over voltage protection. Power to the bath
passes through a filter to prevent switching spikes from being transmitted to
other equipment.
Turn the bath on by switching the control panel power switch to the ON posi-
tion. The stirring motor turns on, the LED display begins to show the bath tem-
perature, and the heater turns on or off until the bath temperature reaches the
programmed set-point. When powered on the control panel display briefly
shows a four digit number. This number indicates the number of times power
has been applied to the bath. Also briefly displayed is data which indicates the
Specific
Specific Heat
Gravity
(cal/g/°C)
1.71 @ 40°C
0.2
0.810 @ 0°C
0.6
0.792 @ 20°C
1.00
1.00
1.05
0.8 @ 0°C
0.87 @ 25°C
0.48 @ 25°C
0.84 @ 75°C
0.53 @ 75°C
0.81 @ 125°C
0.57 @ 125°C
0.92 @ 25°C
0.4
0.934 @ 25°C
0.43 @ 40°C
0.45 @ 100°C
0.482 @ 200°C
0.949 @ 25°C
0.370 @ 40°C
0.393 @ 100°C
0.420 @ 200°C
0.96 @ 25°C
0.4
1.07 @ 25°C
0.358 @ 40°C
0.386 @ 100°C
0.433 @ 200°C
1.11 @ 25°C
0.363 @ 40°C
0.454 @ 100°C
0.505 @ 200°C
0.96 @ 25°C
0.34 @ 100°C
2.0 @ 150°C
0.33
1.9 @ 300°C
1.7 @ 500°C
fr - freeze point
v - viscosity — Flash point test oc = open cup cc = closed cup
8 General Operation
Thermal
Thermal
Conductivity
Expansion
(cal/s/cm/°C)
(cm/cm/°C)
0.0004
0.0011
0.0005 @ 20°C
0.0014 @ 25°C
0.0014
0.0002 @ 25°C
0.001
0.00025 @ 25°C 0.0007 @ 50°C
0.00028 @ 25°C 0.00105
0.00032 @ 25°C 0.00108
0.00034 @ 25°C 0.00107
0.00037 @ 25°C 0.00104
0.00035 @ 25°C 0.00075
0.00035 @ 25°C 0.00077
0.0003
0.00095
0.0014
0.00041
Power
Resistivity
12
Ω-cm )
(10
5 @ 25°C
1000 @ 25°C
10 @ 150°C
1000 @ 25°C
50 @ 150°C
1000 @ 25°C
50 @ 150°C
1000 @ 25°C
50 @ 150°C
100 @ 25°C
1 @ 150°C
100 @ 25°C
1 @ 150°C
100 @ 25°C
1 @ 150°C
1.7 Ω /cm
3
31

Hide quick links: