Emerson Rosemount 8750W Reference Manual page 93

Transmitter with modbus
Hide thumbs Also See for Rosemount 8750W:
Table of Contents

Advertisement

code AX) and the outputs are enabled. If a discrete output is configured for flow limit, the
discrete output will activate when the conditions defined under mode configuration are
met. See
Control
LOI menu path
Modbus coil
This parameter turns the flow limit alert ON or OFF.
ON The transmitter will generate a alert when the defined conditions are met. If a
OFF The transmitter will not generate an alert for the flow limit.
Modbus coil value
0
1
Mode
LOI menu path
Modbus register
The mode parameter sets the conditions under which the flow limit alert will activate. High
and low limits exist for each channel and can be configured independently.
High limit
LOI menu path
Reference manual
Mode
below.
discrete output is configured for flow limit, the discrete output will activate when
the conditions for mode are met.
Flow 1: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 1 > Control 1
Flow 2: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 2 > Control 2
Flow limit 1: 97
Flow limit 2: 98
Configuration
Off (default)
On
Flow 1: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 1 > Mode 1
Flow 2: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 2 > Mode 2
Flow limit 1: 97
Flow limit 2: 98
Flow 1: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 1 > High Limit 1
Flow 2: Detailed Setup > Output Config > DI/DO Config > Flow
Limit 2 > High Limit 2
Advanced Configuration Functionality
89