Troubleshooting
Table 12-12:
Test A. Terminal to front electrode
Test conditions
Location: uninstalled
Required equipment: Multi-
meter
(1)
18 and electrode 18
(1) When the connection head is in the vertical upright position and the flow arrow (see
points to the right, the front of the meter will be facing towards you. Electrode 18 is on the front of the meter. If you cannot determine
the front of the meter, measure both electrodes. One electrode should result in an open reading, while the other electrode should be
less than 0.3 ohm.
Table 12-13:
Test B. Terminal to back electrode
Test conditions
Location: uninstalled
Required equipment: Multi-
meter
(1)
19 and electrode 19
(1) When the connection head is in the vertical upright position and the flow arrow (see
points to the right, the front of the meter will be facing towards you. Electrode 18 is on the front of the meter. If you cannot determine
the front of the meter, measure both electrodes. One electrode should result in an open reading, while the other electrode should be
less than 0.3 ohm.
Table 12-14:
Test C. Terminal to reference electrode
Test conditions
Location: uninstalled
Required equipment: Multi-
meter
17 and process reference
(1)
electrode
(1) Only valid if the sensor has a process reference electrode.
Table 12-15:
Test D. Terminal to case ground
Test conditions
Location: uninstalled
Required equipment: Multi-
meter
17 and safety ground
164
Expected
value
Potential cause
≤ 1 Ω
Shorted electrode
Open electrode
Coated electrode
Expected
value
Potential cause
≤ 1 Ω
Shorted electrode
Open electrode
Coated electrode
Expected
value
Potential cause
≤ 0.3 Ω
Shorted electrode
Open electrode
Coated electrode
Expected
value
Potential cause
≤ 0.3 Ω
Moisture in terminal block
Leaky electrode
Process behind liner
®
Rosemount
8750W Transmitter with Modbus Protocol Reference Manual
Corrective action
Replace sensor
Remove coating from sensor
wall
Section
3.2.3) on the connection head flange
Corrective action
Replace sensor
Remove coating from sensor
wall
Section
3.2.3) on the connection head flange
Corrective action
Replace sensor
Remove coating from sensor
wall
Corrective action
Clean terminal block
Replace terminal block
Replace sensor