Pre-use checkout

Advertisement

1
10 LPM
Off
3
2
"88" Cylinder Duration
5 L/min
10 L/min
5
39 Days
19.5 Days
10
19.4 Days
9.7 Days
20
9.6 Days
4.8 Days
40
4.7 Days
2.3 Days
80
2.2 Days
1.1 Days
* All calculations in the table above are based on a full
cylinder, 138 bar (2000 psig), 1963 liter "88" cylinder, with
a cylinder change at 14 bar (200 psig). The figures are
calculated based on a total continuous breathing circuit
gas flow and a cylinder conversion factor of 14.2 liters per
bar/0.98 liters per psig.
• INOMAX flow = [Desired dose × total ventilator flow] ÷
[Cylinder concentration - desired dose]
• Cylinder volume = Cylinder conversion factor × cylinder
pressure (bar/psig)
• Cylinder duration (hours) = (Cylinder volume ÷ INOMAX
flow rate) ÷ 60
Mallinckrodt Manufacturing LLC, 6603 Femrite Drive, Madison, WI 53718-6801 USA
For assistance contact Technical Support 877-566-9466.
Pre-Use Assembly
4
5
6
10
9
FLOW
20 L/min
40 L/min
9.8 Days
4.9 Days
4.8 Days
2.4 Days
2.4 Days
1.2 Days
1.2 Days
14 Hours
13.3 Hours 6.6 Hours
1. O
2. O
3. 15M x 4.5 mm Adapter
4. 22M / 15F x 22M / 15F Adapter
5. Injector Module
7
6. 300 mm of 22 mm hose
7. Patient Gas Sample Line with Nafion
8. Gas Sample Tee
9. Injector Module Electrical Cable
10. NO/N
8
FiO
Dilution
2
10
20
!
40
!
80
!
Flowmeter
2
Tubing
2
Injector Tube
2
Set FiO
2
.21
.40
.60
0.21
0.40
0.59
0.20
0.39
0.59
0.20
0.38
0.57
0.19
0.36
0.54
Actual FiO
.80
1.00
0.79
0.99
0.78
0.98
0.76
0.95
0.72
0.90
2
20753 Rev 01
2014-08